Service
Regular servicing by a Sartorius techni-
cian will extend the service life of your
moisture analyzer and ensure its contin-
ued weighing accuracy. Sartorius can
offer you service contracts, with your
choice of regular maintenance intervals
ranging from 1 month to 2 years.
The frequency of the maintenance
intervals depends on operating condi-
tions and user tolerance requirements.
Repairs
Repair work must be performed by
trained service technicians. Any
attempt by untrained persons to per-
form repairs may lead to hazards
for the user.
Cleaning
!
Make sure that no dust or liquid enters
the moisture analyzer housing
!
Do not use any aggressive cleaning
agents (solvents, abrasive cleaning
agents, etc.); clean the moisture analyz-
er using a piece of cloth which has been
wet with a mild detergent (soap) only
§
Unplug the power cord from the wall
outlet (mains supply)
If you have a cable connected to the
interface, unplug it from the moisture
analyzer
$
The pan draft shield and the pan
support can be removed for cleaning
§
Carefully remove any sample residue/
spilled powder using a brush or a hand-
held vacuum cleaner
§
After cleaning, wipe down the analyzer
with a soft, dry cloth
Cleaning the Heating Unit and
Temperature Sensor
§
Open the hinged cover
!
Danger:
The terminals of the heating
unit are under live current
§
To disconnect from AC power,
unplug the power cord from the wall
outlet (mains).
If you have a cable connected to the
interface port, disconnect it from the
moisture analyzer
§
Carefully remove any residue from the
temperature sensor
§
Use a soft-bristled brush or a commer-
cially available cleaning agent for
glassware to clean the ceramic heating
element
46
Care and Maintenance