Example: Standard Drying with Fully Automatic Shutoff Mode
The moisture content of 2 g of corn starch is to be analyzed. The sample can scorch if overheated, but it is not particularly heat sensitive.
The analysis is to be ended automatically as soon as a constant weight is reached.
Settings (changes in the factory settings required for this example):
Program number:
1
Final temperature:
130°C
Start of analysis:
With stability after “Enter” key is pressed
End of analysis:
automatic (factory setting)
Part A: Configuring the Drying Parameters
Step
Key (or Instruction)
Display/Printout
1. Turn on the moisture analyzer
e
Sartorius logo is displayed
Self-test runs
2. Select the
PROGRAM
function
y
or
x
3. Confirm the
PROGRAM
function
u
4. Select
P1
program
u
5. Select
Heating
program
u
6. Select
Standard drying
u
7. Enter the temperature:
130
u
3 ƒ
x
u
5 ƒ
y
u
27
PROGRAM
P1 105
o
C Auto.
S
0
092
g
TARE
SETUP