Setting the Drying Parameters
Purpose
Adapt the moisture analyzer to the special requirements of products.
Parameters can be configured individually for each program.
Drying Parameters (Overview)
ο
Factory setting
√
User-defined setting
Program
ο
P1
memory
P2
(
PROGRAM
)
P3
Heating program
ο
Standard drying
Temperature
105°C
40...230°C
MA30 mode
Temperature
105°C
40...160°C
1
)
Standby
ο
Off
temperature
2
)
On
Temperature
40°C
40...100°C
Initial weight
ο
Off
On
Target
5.0 g
0.2 g...35.0 g
Start analysis
With stability
when Enter pressed
ο
Without stability
when cover is closed
End of analysis
ο
Fully automatic
Semi-automatic
Loss/24 s
5 mg
1...20 mg
Timer mode
Time
15.0 min
0.1...99.9 minutes
Display mode
ο
Moisture (%L)
Dry weight (%R)
Ratio (%LR)
Residual weight (g)
Residual weight (g/kg)
ο
Off
intermediate results
On
Interval
0.1 min
0.1...10.0 minutes
Factory settings
No
Yes
1
) Marked in Info/Status line by an asterisk (“
*
”); e.g. “
*105
o
C
”
2
) Not when using the “MA30 mode” heating program
24
Operating the Moisture Analyzer