5.2
pH - Calibration
The pH electrode is calibrated by a two-point calibration using liquid buffers. The calibration is done
before the electrode is installed in the culture vessel. For details about preparing of the pH-electrode
and handling during calibration see the operating information delivered by the manufacturer. For de-
tails about mounting the electrode into the culture vessel, see the fermentor´s operating manual.
The calibration determines the „zero drift“ and „slope“ of the electrode. The measured pH-value is cal-
culated from the electrode voltage using the Nernst equation, considering the „zero drift“, „slope“ and
„temperature“. During calibration the temperature is entered manually. For pH-measurement during
the process temperature effects are compensated via the measured temperature in the fermentor.
The impact of heat during the sterilization can cause a zero drift of the electrode. To compensate for
such effects the DCU3 - system allows to recalibrate the pH-electrode during the process. You can
measure the pH-value in a sample taken from the fermentor and enter this value into a corresponding
menu. The software recalculates the zero drift and displays the corrected „in-situ“ pH-value afterwards.
The operating display for the pH - electrode shows the pH value, the electrode voltage and the elec-
trode parameters „zero drift“ and „slope“. Thus the functioning of the electrode can be easily checked.
5.2.1 Operating
Display
Tag
Entry Displayed value or function, required input
pH
pH
Display of measured pH value,
input of pH-value measured in sample for recalibration
ELEC
mV
Display of measured electrode voltage (raw signal)
TCOMP
Display “temperature compensation”
TEMP
Display of temperature for temperature compensation
ZERO
mV
Display of zero drift
SLOPE
mV/pH Display of slope
Start Calibration
Enabling the calibration function
Start Recalibration
Enabling the recalibration function (during the process)
Содержание DCU 3
Страница 2: ......
Страница 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Страница 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Страница 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Страница 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Страница 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Страница 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Страница 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Страница 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Страница 114: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Страница 139: ......
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......
Страница 148: ......
Страница 149: ......
Страница 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...