Operators Handbook DCU3 – System
14., Supplement
BAEDCU3 - Rev. 4.6 - 0504
14 - 1
14 SUPPLEMENT
14.1
Alarms, Meanings and Remedial Measures (Troubleshooting)
14.1.1 Process
Alarms
The operator can individually switch-on and off the alarms referred to in the table below :
Text of alarm line
Meaning
Remedial measures
„Name State alarm ...“
Alarm of digital input „name“
confirm alarm with 'ACK'nowledge
„Name low alarm ...“
the process value „name“ is below its
lower limit threshold
confirm alarm with 'ACK'nowledge
„Name high alarm ...“
the process value „name“ is below its
upper limit threshold
confirm alarm with 'ACK'nowledge
„Fuse blown ...“
automatic cut-out of the supply unit is
activated
switch on automatic cut-out again; if
alarm is repeated contact service
„Heater failure ...“
overheat protection of thermostat cy-
cle is activated
thermostat cycle must be refilled
(see details in Operating Manual of
fermentor)
„Motor failure ...“
overheat protection of motor is acti-
vated
let motor cool down before switch-
ing on fermentor again
„TEMP Sens. failure ...“
Pt-100 in culture vessel is defective or
cable not plugged
check cable /plug, if connection is
ok, sensor may be defective, then
replace sensor
„JTEMP Sens. failure ...“ Pt-100 in thermostat cycle is defective contact BBI´s customer service
„TURB Sens. failure ...“
sensor turbidimetry measurement dis-
connected or defective
check sensor; replace, if necessary
14.1.2 Process
Messages
Text
Meaning
Measures
„Steriliz. Finished ...“
Sterilisation is finished
After confirmation with 'ACK'nowl-
edge the fermentation process can
be started
„Shut down DCU ...“
Emergency cut off - key of DCU en-
abled
Release 'SHUT DOWN' state by
pressing the key again
„Shut down fermenter
...“
Emergency cut off - key of fermentor
enabled
Switch-on fermentor by pressing
the SHUT DOWN key again
„Name <3mA alarm ...“
„(Name of an analog in-
put) < 3 mA alarm“
The analog 4..20 mA input has a bro-
ken cable
Check cable connection and con-
nected peripheral unit
Содержание DCU 3
Страница 2: ......
Страница 4: ...Operators Handbook DCU3 System Introduction BAEDCU3 Rev 4 6 0504...
Страница 8: ...Operators Handbook DCU3 System Contents BAEDCU3 Rev 4 6 0504...
Страница 10: ...Operators Handbook DCU3 System 1 Design and Function BAEDCU3 Rev 4 6 0504 1 2...
Страница 74: ...Operators Handbook DCU3 System 6 Main function Control Loops BAEDCU3 Rev 4 6 0504 6 40...
Страница 80: ...Operators Handbook DCU3 System 7 Main function Batch Control BAEDCU3 Rev 4 6 0504 7 6...
Страница 88: ...Operators Handbook DCU3 System 8 Main Function Recipes BAEDCU3 Rev 4 6 0504 8 8...
Страница 106: ...Operators Handbook DCU3 System 9 Main function Maintenance BAEDCU3 Rev 4 6 0504 9 18...
Страница 110: ...Operators Handbook DCU3 System 10 Main Function Documentation BAEDCU3 Rev 4 6 0504 10 4...
Страница 114: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...
Страница 139: ......
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......
Страница 148: ......
Страница 149: ......
Страница 150: ...Operators Handbook DCU3 System 14 Supplement BAEDCU3 Rev 4 6 0504...