Malfunctions
10 Malfunctions
10.1 Status Messages
Status
Message
Fault
Cause
Remedy
Chapter,
page
OLDER?
If a USB mass storage
device with a software
update is connected:
The device detects an
obsolete update
version.
The software update
version is older than the
software version on the
device.
To perform the update
anyway: Confirm the display.
LOCKED
A software update is
to be installed on a
conformity-assessed
device.
Software updates are
not
permitted for conformity-
assessed devices.
If the software update has
received corresponding
approval: The software
update can be uploaded
into the device by Sartorius
Service. Re-verification may
then be necessary.
INVALID.
If a USB mass storage
device with a software
update is connected:
The device cannot
find a valid update.
No valid manufacturer
software updates exist on
the connected USB mass
storage device.
Check whether files exist on
the USB mass storage
device. Procure the software
update files from Sartorius
Service.
10.2 Warning Messages
Warning
Message
Fault
Cause
Remedy
Chapter,
page
APP.ERR.
The device has
measured an invalid
weight value.
The applied weight is
too low.
Increase the applied weight
to more than the minimum
load.
The weight value is
negative.
No
sample has been placed
on the balance.
Place the sample on the
balance.
DIS.ERR.
The value to be
output
cannot
be
shown in the
operating display.
The data to be displayed is
not
compatible with the set
display format.
Adjust the display settings
in the menu, e.g. resolution,
unit, decimal places.
HIGH
The device is
overloaded.
The device’s maximum
weighing capacity has been
exceeded
Reduce the applied weight
to below the device’s
maximum weighing
capacity.
LEVEL.ERR.
The device is
not
measuring a weight
value.
The device is
not
leveled.
Level the device.