Operating Instructions Cubis
®
MCE
47
Operation
8 Operation
8.1
Switching the Device On and Off
The device only delivers accurate values if it has reached the necessary
operating temperature. The warm-up time after switching the device on
must therefore be complied with.
Procedure
t
Connect the device to the power supply.
t
If the device does not switch on automatically after it is connected to
the power supply: Press the on key on the electronics module.
t
To allow the device to reach the necessary operating temperature to
accurately determine the weight: Comply with the warm-up time of
2 hours after switching the device on.
y
If this relates to a conformity-assessed device: The weight value is
marked as invalid during the warm-up period.
t
NOTICE
Damage to the device when disconnecting the weighing
module and electronics module! The connection of the weighing
module and the electronics module must not be disconnected during
operation.
t
Level and adjust the device.
y
The device is ready for use.
t
NOTICE
Pointed or sharp-edged objects may damage the operating
display! Touching the operating display with objects, such as pens, may
damage the touchscreen surface of the operating display. Only touch
the operating display with your fingertips.
t
To switch the device off: Press and hold the [On | Off] key.
8.2 Leveling the Device
8.2.1 Leveling the Device with the Level
Leveling compensates any inclines at the device’s installation site. Minor
inclines can be compensated with the front leveling feet. If leveling is
necessary: The level’s air bubble is not located within the circle marking.