58
Operating Instructions Cubis
®
MCE
Cleaning and Maintenance
9 Cleaning and Maintenance
9.1
Preparing the Device
Procedure
t
Turn the device off.
t
Disconnect the device from the power supply. To do so, disconnect the
power supply cable from the wall outlet.
t
Remove weighing pan and all associated components from the weighing
module, e.g. shield frame, pan support.
9.2 Cleaning the Device
NOTICE
Corrosion or damage to the device due to unsuitable cleaning agents!
t
Do
not
use corrosive, chloride-containing, or aggressive cleaning agents.
t
Do
not
use cleaning agents that contain abrasive ingredients, e.g.
scouring agents, steel wool.
t
Do
not
use solvent-based cleaning agents.
t
Check whether the cleaning agents used are compliant materials (see
Chapter “15.4 Materials”, page 69).
t
Make sure that cleaning materials, such as cloths, are only slightly damp.
Procedure
t
NOTICE
Malfunction or damage to the device due to the ingress of
moisture or dust!
t
Remove dust and powdery sample residue with a brush or hand-held
vacuum cleaner.
t
NOTICE
Corrosion or damage to the weighing pan due to unsuitable
cleaning agents!
t
Wipe down the weighing pan and the shield plate with a cleaning
agent and a cloth. The cleaning agent must be suitable for stainless
steel and titanium.
t
Wipe the associated components of the weighing pan with a brush or
a damp cleaning cloth, e.g. shield plate, support pan.
t
Wipe the device housing with a damp cloth. For more severe
contamination, use a mild soap solution.
Tip
We recommend cleaning the weighing pan on a regular basis, e.g. weekly.
Do not allow deposits to form on the weighing pan.
Protective oil may be applied to the weighing pan for additional protection.
The protective oil must be suitable for stainless steel and titanium.
Содержание Cubis MCE11201S
Страница 2: ......
Страница 78: ...78 Operating Instructions Cubis MCE Conformity...