Operating Design
22 Arium
®
Smart Station Operating Instructions
4.5 Message List
All active messages can be viewed in the message list. The cause of the
message does not necessarily lie with the dispensing unit, but usually with
the connected Arium
®
water treatment device.
The messages in the message list are sorted according to priority. Error
messages appear first in the list. Within the same priority level, the messages
are sorted according to date and time.
Messages
cannot
be deleted manually. They remain in the message list and
will be displayed on the operating display until their cause has been
corrected. The device detects if the cause of a message has been corrected
and automatically deletes the message from the message list.
Error
Error 0130
Care
Change final filter
Care
Change cartridges
2/1/2021
11:46
2/3/ 2021
9:33
2/3/2021
14:17
Messages
1
2
4
3
Fig. 13: Message list (example)
Pos .
Symbol
Name
Description
1
8
Message
symbol
Displays an error message.
Displays a warning message.
Displays a status message (info).
2
Occurrence of
the message
Displays the date and time when the
message occurred.
3
Brief
description
Displays the error number or a short
description of the message.
4
Message type
−
Error: Displays and opens an
error message.
−
Warning category, e.g. care:
Displays and opens a warning
message.
−
Service: Displays and opens a
status message.
Содержание Arium H2O-ARST-P-B