6. Saving, retrieving and deleting results
• The results of each successful measurement are stored
together with date and time. With more than 30 items of
measured data, the earliest items of data measured are lost.
• In order to retrieve measurement results again, press the
memory button “M”. Use the SET button to select the
required user memory and press the memory button “M”.
Initially, the average value of all stored results in this user
memory is first displayed (display “A”).
If you continue pressing the memory button “M”, the last
results in each case are measured are displayed with date,
time and graphic WHO classification.
With the individual stored results, warning limits and indi-
vidual stored results are also displayed alternately, without
pressing any other buttons.
• After pressing the memory button “M” and selecting the
user, you can clear the memory by holding down the
MODE and SET buttons simultaneously for 5 seconds.
7. Error messages/troubleshooting
In case of faults, the
message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to
, the cardiac arrhythmia icon
appears in the
display),
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that you
do not move or talk. Re-insert the batteries if necessary, or
else replace them.
8. Cleaning and Care
• Clean your blood pressure computer carefully using only a
slightly moistened cloth.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.
9. Technical details
Model no.
SBM 09
Measurement
method
Oscillating, non-invasive blood
pressure measurement on the wrist
Measurement range Pressure 0 – 300 mmHg,
heart rate 40 – 199 beats/minute
Precision of pressure
reading
Systolic ± 3 mmHg / diastolic ± 3 mmHg
pulse ± 5 % of displayed value
Measurement uncer-
tainty
Max. admissible standard deviation
according to clinical examination.
systolic 8 mmHg / diastolic 8 mmHg
Memory
3 x 30 storable data lines
Dimensions
L84 x W74 x H35 mm
Weight
Approx. 140 g
Cuff size
For a wrist circumference of approx.
135 to 195 mm
Admissible operating
temperature
+10 °C to + 40 °C,
< 85 % relative humidity