Note: The alarm output port is connected with a set of normally open and normally closed
contacts of the relay inside the machine. The working current should be smaller than 3A while
the working voltage should be smaller than 300V.
VII
.
Maintenance
The equipment must be shut down for maintenance, and the power supply must be cut
off. The operation can only be carried out after 3 minutes, otherwise there will be a risk of
electric shock. When the room temperature is lower than 2°C, the internal water must be
drained when the machine is shut down for a long time.
Regularly disassemble the filter gauze for cleaning and use a compressed air gun to blow
away the dust on the condenser (about 1 week is suggested and the air filter must not be
missing for a long time).
VIII. Simple Troubleshooting
Failure
Failure Cause
Approach
Machine turned on
but unelectrified
Power cord is not plugged in place
Check and ensure the power
interface and the power plug is
plugged in place and in good
contact.
Fuse burnt-out
Pull out the fuse box from the
power supply interface of the
chiller, check the fuse, replace
Display
Condition
Alarm
code
Buzzer
OUT H1
、
H2
OUT H1
、
H3
Circulating pump works properly
DISCONNECTION
BREAKOVER
Ultrahigh room temp
E1
Sounds
BREAKOVER
DISCONNECTION
Ultrahigh water temp
E2
Sounds
BREAKOVER
DISCONNECTION
Ultralow water temp
E3
Sounds
BREAKOVER
DISCONNECTION
Faulted room temp sensor
E4
Sounds
BREAKOVER
DISCONNECTION
Faulted water temp sensor
E5
Sounds
BREAKOVER
DISCONNECTION
Flow rate 1 alarm
E6
Sounds
BREAKOVER
DISCONNECTION
Flow rate 2 alarm
E7
Sounds
BREAKOVER
DISCONNECTION
Chiller power failure
BREAKOVER
DISCONNECTION