TY-JS-220308-001

12

don’t release the buttons until the controller displays rE. The con troller works in normal

order in 6s and is restored to factory set tings.

3.Press

key to show t1. Press

key again to show room

temperature (Meanwhile D1 is flashing, displaying room temperature)

Continue to press

key to show t2. Press

key again to show flow rate 1.(Unit:

L/min)

Continue to press

key to show t3. Press

key again to show flow rate 2.(Unit:

L/min)

Continue to press

key to exit display or restore to water temperature display in 6

seconds if no key is pressed.

4. Alarm function

(1) Alarm Display:

When alarm occurs, the error code and the temperature will be alternately displayed.

E1

E2

E3

E4

E5

E6

E7

Ultrahigh

room

temperature

Ultrahigh

water

temperature

Ultralow

water

temperature

Room

temperature

sensor

failure

Water

temperature

sensor

failure

Flow

rate

1

alarm

Flow

rate

2

alarm

(2) To suspend the alarm:

In alarming state, the alarm sound could be suspended by pressing any button, but

the alarm display remains until the alarm condition is eliminated.

5. General settings adjustment

Press SET button

SET

to enter into the user-defined state. Meanwhile, D1 flashes

to indicate that the controller is in parameters setup status.

(1)Under intelligent mode, the control panel displays the temperature difference value

between water and room (Factory setting is -2).

(2)At this moment, press ▲▼ key to change settings. After modifying the value, press the

ENTER button (RST) to save and exit, then new parameters take effect, or press SET

key (SET) to exit without saving parameters. If there is no more action within 20

seconds, it will automatically exit modifying status without saving parameters.

6. User parameter adjustment

Press and hold the ▲ key first, and then press "SET" key at the same time for 5 seconds

until 00 is displayed, at this time, press the ▲▼ key to select the set password (F7=8),

and then press "SET" key. If the password is correct, F0 is displayed, entering the setting

state. The panel indicator D1 flashes indicating the controller being in the parameter

setting state. If the password is wrong, it will return to temperature display. After entering

the setting state, Press the ▲ key to cycle into the selected items, and press ▼ key to

cycle in the opposite direction. (Only F0-F11 is displayed) Select the set item and then

press the "SET" key to enter the next parameter modification, and the original set value

is displayed. At this time, press ▲▼ key to modify the parameter value, and then press

"SET" key to return to the previous menu setting item. Press the "RST" key at any time to

save the modified parameters and exit, return to the temperature display, and run

according to the new parameters. If no key is pressed within 20 seconds, the controller

will automatically exit the parameter setting state without saving the modified

parameters. (In the parameter setting state, the system runs according to the parameters

adjusted immediately.)

7. Advanced parameters adjustment case

Содержание CW-5000 Series

Страница 1: ...TY JS 220308 001 1 CW 5000 5200 Industrial Chiller User Manual...

Страница 2: ...rly so that you can refer to it whenever you need information This manual is not a quality guarantee Our company reserves the right to the interpretation of the correction of misprint and improperly d...

Страница 3: ...NTENTS Notice 4 I Overview 7 II Model Illustration 7 III Product Outlines Parts 8 IV Installation 9 V Operating States Parameters Adjustment 11 VI Alarm Output Terminal 13 VII Maintenance 14 VIII Simp...

Страница 4: ...on a The power supply must conform to the standard indicated in the nameplate or the manual b Use the standard cable and select the standard wire diameter c The grounding wire must be installed and th...

Страница 5: ...ailable Otherwise it may cause anesthesia and suffocation if a large amount of refrigerant fills the enclosed space b Avoid contacting with skin or it will cause frostbite DANGER Failure to take safet...

Страница 6: ...r filter at least once a week If it is blocked the cooling ability will decrease and the power consumption will increase and the alarm will not work properly Equipment surface cleaning a Please use cl...

Страница 7: ...chiller the refrigerant in the evaporator coil absorbs the heat of the return water and vaporizes into steam The compressor continuously extracts the generated steam from the evaporator and compresses...

Страница 8: ...rt names are only for one of the models and they will be slightly different in different models POWER SWITCH TEMPERATURE CONTROLLER AIR INLET FILTER GAUZE ALARM OUTPUT TERMINAL COOLING WATER OUTLET CO...

Страница 9: ...Requirements Cooling medium allowed by the chiller includes purified water distilled water high purity water and other softened water It is forbidden to use oily liquids liquids containing solid part...

Страница 10: ...on 1 Please select the cable based on the max rated current indicated on the label of the chiller 2 Recommended reference standard for power cord diameter selection Rated current A 5 10 15 25 35 50 Wi...

Страница 11: ...trolling parameters according to room temperature for meeting equipment cooling requirements The new T 504D intelligent temperature controller is selected constant temperature control mode as factory...

Страница 12: ...ey to change settings After modifying the value press the ENTER button RST to save and exit then new parameters take effect or press SET key SET to exit without saving parameters If there is no more a...

Страница 13: ...the ambient the fluctuation will not exceed 0 5 There will be an alert when water temperature is 10 lower or higher than target e g when ambient temperature is 30 0 cooling water temperature is betwe...

Страница 14: ...the air filter must not be missing for a long time VIII Simple Troubleshooting Failure Failure Cause Approach Machine turned on but unelectrified Power cord is not plugged in place Check and ensure th...

Страница 15: ...tlet and inlet To ensure a smooth ventilation for air outlet and inlet Voltage is extremely low or astable To improve the power supply circuit or use a voltage regulator Improper parameter settings on...

Отзывы: