Operation of Media 7 Differential Pressure Meter
1. Press key (in start screen) to go to the main menu.
2. Select Start-up (1) with or
key and confirm with
key.
3. Select User level (1.1) with or
key and confirm with
key.
4. Press key and select 'Specialist' with or key.
5. Press
key to confirm the setting.
Activating password protection
The password can only be changed and activated in the Specialist user level.
1. Press key (in start screen) to go to the main menu.
2. Select Start-up (1) with or
key and confirm with
key.
3. Select Password protection (1.7) with or key.
4. Press key to activate password protection.
Î
The default password is
1234
. To change the password, proceed as follows:
1. In the Start-up (1) menu, select Password (1.8) with or
key and confirm with
key.
2. Press key. Select the digit within the password with or key.
3. Press key. Change the number (0 to 9) within the password with or key.
4.
Confirm with
key (proceed in the same way for the rest of the password).
5. After selecting all digits of the password, press key.
8.1.4
Performing zero calibration
We recommend performing a zero calibration after first start-up and after changes have
been made to the plant.
To set zero, the pressure in the measuring lines must be equal. To check zero, make sure that
the pressures in both measuring chambers are identical at atmospheric pressure, i.e. the sig-
nal at the terminals 31 and 32 must be 4 mA at a differential pressure of ∆p = 0 mbar (see
Fig. 7 on page 37).
At ∆p = 0 mbar, the 0.0 % reading must be displayed.