4.1 Display and operating elements

All the necessary information and measured
values stored in the instrument’s memory
are shown in the display of 

Media 6 with

LCD

. Although the 

Media 6 with LED

 ver-

sion does not have such a digital display,
important operating conditions are indi-
cated by the LED. 

Three keys are used to operate the differen-
tial pressure meter:

Up key
Down key
Enter key

and a DIL switch with four switches to select
the type of gas and write protection function.

4.1.1 Changing over display mode
for Media 6 with LCD

Press the   key to change over from the
standard display to seven other display
modes.
After 8 seconds or after the message has fin-
ished running across the display, the dis-
play returns automatically to the standard
display. 

O2

e.g. gas name and current tank 
content

P

current differential pressure

MCN

maximum capacity nominal

MCN/R

100 % capacity assigned to 
20 mA signal

SCN

save capacity nominal
geometric capacity up to overflow/
gauge pipe

SCN/R

100 % capacity assigned to 
20 mA signal

UCW

useable capacity work 

P100

maximum differential pressure

PTANK

nominal tank pressure  

Indicated value corresponds to the 
pressure assigned to the density

(liquid)

as per vapor pressure diagram. 
If the calculations for 

MCN

 and

 SCN

 

are based on density

(liquid)

 at 1 bar 

abs, then 1 bar is shown for 

PTANK

.

X-TANK-16 

e.g. tank ID as running text

ERROR

error message displayed auto-
matically when an error occurs
(see section 8 on troubleshooting).

OFF

Special signal on opening the
equalizing valve, I = 3.6 mA
(refer to EB 9527-2, page 7).

EB 9527-1 EN

15

Operation

Содержание 5006-0

Страница 1: ...sure Meter Media 6 with LCD Media 6 with LED Mounting and Operating Instructions EB 9527 1 EN Firmware A 2 11 LCD B 2 11 LED Edition July 2004 Fig 1 Media 6 with LCD and attached valve block Media 6 w...

Страница 2: ...isplay mode for Media 6 with LCD 15 5 Start up 16 6 Setting 16 6 1 Write protection 16 6 2 Selecting the gas type 16 6 3 Checking zero 17 6 4 Checking measuring range span 19 6 5 Setting limit switche...

Страница 3: ...are authorized to work on explosion protected devices in hazardous areas See section 9 for more details Any hazards which could be caused by the process medium and the oper ating pressure in the instr...

Страница 4: ...figured over software as minimum and maximum alarms They can be adjusted separately over the keys on the device Filling limit during oper ation The filling limit during operation UCW can be set over t...

Страница 5: ...Previous version New version EB 9527 1 EN 5...

Страница 6: ...ay 7 LCD or LED and the D A con verter 9 for the output signal which is is sued at connector A as a two wire 4 to 20 mA transmitter signal The SERIAL INTERFACE 10 enables the device to be configured u...

Страница 7: ...ndicating unit with LCD or LED 1 Differential pressure cell 1 1 Measuring diaphragm 1 2 Measuring springs 1 3 Lever 1 4 Elastic disc 1 5 Diaphragm axis 2 Displacement sensor 3 Microprocessor 4 Memory...

Страница 8: ...static pressure ambient temperature 0 03 1 bar at 20 to 70 C on zero and span 0 25 10 K Limit switches 2 software limit switches A1 and A2 configurable as minimum or maximum alarms in accordance with...

Страница 9: ...request Measuring springs CrNi steel Diaphragm plates and function parts Lever Indicating unit Polycarbonate polyamide Note Devices intended for oxygen service are labeled Oxygen Keep free of oil and...

Страница 10: ...with clamp For wall mounting use a mount ing part without clamp See the dimen sional drawing on page 29 for mounting to control panels Note We recommend installing one shut off valve in each measurin...

Страница 11: ...ew fittings and SAMSON valve blocks are listed together with their order numbers in the Data Sheet T 9555 EN Shut off valve Fig 4 SAMSON valve block Test connection Equalizing valve lead sealable Shut...

Страница 12: ...connection of intrinsically safe circuits For intrinsically safe electrical ap paratus that are certified according to the Directive 94 9 EC the data specified in the EC type examination certificate a...

Страница 13: ...st ter minals and for checking the output signal on calibration The output signal of the two wire circuit is not interrupted Make sure that there is a load of 0 4 V DC at the ammeter for the test conn...

Страница 14: ...y and p Nom perm capacity SCN Tank capacity Gas type and operating status Factor Tank designation Light emitting diode Plug in plate for gas types and measuring ranges Note A quick guide is located be...

Страница 15: ...the display the dis play returns automatically to the standard display O2 e g gas name and current tank content P current differential pressure MCN maximum capacity nominal MCN R 100 capacity assigne...

Страница 16: ...data from being changed unintentionally SPAN PROTECTION as an additional write protection for the span setting The write protection at switch 4 of the DIL switch must first be switched OFF before var...

Страница 17: ...spheric pressure At a differential pressure of p 0 mbar the current signal at connector A or at the TEST connection must be 4 mA See test arrangement in Fig 9 Note When activating the gas column corre...

Страница 18: ...X mbar appear in the display Current signal I indicates the present mA value Press key Zero is adjusted Release key 0 mbar appears in the display Current signal I 4 mA corresponding to the liquid lev...

Страница 19: ...e write protection turn switch 4 ON 6 4 Checking measuring range span A basic calibration with a linear charac teristic based on the upper range value of the dp cell has been made at the factory de fa...

Страница 20: ...ed down the current measured value is shown in the display Current signal I shows the present mA value Press key the span is calibrated current signal goes to 20 mA Readout p 100 Release key Turn writ...

Страница 21: ...edia 6 with LCD Alarm A1 and A2 markers Both limit switches are already set over the software either as min or max contacts A1MIN or A1MAX as well as A2MIN or A2MAX appear in the display Both limit sw...

Страница 22: ...owly Press key to display the currently set A1 alarm at the ammeter The LED lights up Press to reduce the value in steps of 1 or Press to increase the value Press key to confirm new setting The LED go...

Страница 23: ...and 22 8 mA Release key current signal I indicates the mA value corresponding to the tank capacity Write protection turn switch 4 ON Media 6 with LED Write protection turn switch 4 OFF 4 mA ammeter Pr...

Страница 24: ...ated with ear lier TROVIS VIEW software versions 1 02 to 2 20 are not 1 1 compatible with Media 6 devices with firmware versions A 2 10 or B 2 10 They must first be loaded and converted over TROVIS VI...

Страница 25: ...ata is saved in the memory pen the memory pen can be removed Media 6 with LED Data transfer from Media 6 to memory pen write and from memory pen to Media 6 read Status write and read Insert memory pen...

Страница 26: ...th LED Thus error code 1 is signaled as follows long medium 9x short Refer to the table on page 27 for the de scription of the error codes Reset or acknowledge errors by pressing the key Any new error...

Страница 27: ...ion produces values that exceed the permissible limits These values are not saved in EEPROM Check zero and span calibration On doing so pay attention to the p If the fault remains send device to SAMSO...

Страница 28: ...mponent of the Media 6 on which the explosion protection is based must be serviced the instrument must not be put back into operation again until an expert has inspected the device accord ing to explo...

Страница 29: ...ring chamber rear side to fit M8 screws Two bore holes 8 5 mm for attachment to the valve block Control panel Pressure gauge acc to manufacturer Pressure sensor acc to manufacturer Pressure gauge conn...

Страница 30: ...30 EB 9527 1 EN...

Страница 31: ...EB 9527 1 EN 31...

Страница 32: ...32 EB 9527 1 EN...

Страница 33: ...EB 9527 1 EN 33...

Страница 34: ...SAMSON AG MESS UND REGELTECHNIK Weism llerstra e 3 60314 Frankfurt am Main Germany Phone 49 69 4009 0 Fax 49 69 4009 1507 Internet http www samson de S Z 2004 10 EB 9527 1 EN...

Отзывы: