Operating Instructions
58 / 92
5
Calibrate
touch
Touch
auto zero
Calibrating/aligning the display
Caution!
An incorrect calibration will result in an inability to work
the buttons by tapping them! In that case, the calibration process
must be carried out with a mouse! Connect the PC mouse via
the USB connection and follow the on-screen instructions using
the cursor. The mouse connected is only recognized after the
appliance is restarted (USB connection optional).
.
Tap the key "Touch calibration".
D
A window with a crosshair in the centre is displayed.
D
Important, only tap the crosshairs. Do not touch the
clear area, otherwise malfunctions occur.
.
Tap the crosshair.
D
The crosshair switches to the lower right corner.
.
Tap the crosshair.
D
The crosshair switches to the next corner.
.
Tap the crosshair until all corners have been con
firmed.
D
A note on the save process appears.
.
Tap the key "Confirm".
D
The note disappears.
D
Calibration is completed.
.
Continue to the next window - tap arrow key.