The fiber cable should never be disconnected. The boots in the cables protect the connectors
from movement, shock and dust.
There are no serviceable parts inside the probe. Return the probe in case probe fails to meet
its specs or does not operate. Opening the probe voids the warranty.
EQUIPMENT CODES
PERFORMANCE VERIFICATION
XFVP-L probes are shipped from Saker and both probe head and receiver are tested as a
system and adjusted to meet the DC gain, offset and bandwidth specification.
Equipments needed to perform verification:
-
Bench Multimeter of at least 4.5 digits
-
Adjustable power supply or precision voltage source
-
Banana to BNC cable (Pomona 4531 or similar)
-
BNC-SMB adapter (Amphenol 242185 or similar)
-
BNC-BNC cable (Caltest CT2942 or similar)
-
50 Ohm terminator (Pomona 3285 and Amphenol 202104)
-
BNC to banana Jack adapter (Caltest CT2939 or similar)
CODE
Description
XFVP-L10-x
XFVP-L probe 110MHz with 0.5m BNC output cable, SMB
unterminated coax cable, 2xAA batteries, tripod, adjusting tool, quick
guide, carrying case, 1m USB cable. Specify -x for length, standard
2m
XFVP-L20-x
XFVP-L probe 220MHz with 0.5m BNC output cable, SMB
unterminated coax cable, 2xAA batteries, tripod, adjusting tool, quick
guide, carrying case, 1m USB cable. Specify -x for length, standard
2m
XFVP MODEL
XFVP SERIAL NUMBER #
DATE
BR-XFVPL-Rev 2. 12/22