Set the oscilloscope vertical scale at an appropriate scale given the conversion ratio of the
XFVP-L and attenuator (see specifications section) in question. A 50 termination is necessary
for calibration and signal integrity.
The preferred method of operation is as follows:
1. Place and secure the probe head
2. Connect the probe head to the DUT using the appropriate attenuator.
3. Connect the receiver to the oscilloscope
4. Turn on the probe head and receiver.
5. Press the “Offset Cal” key in the receiver. Power on the DUT.
6. Take measurements as needed.
OPERATION
TURNING ON
Ω
BR-XFVPL-Rev 2. 12/22
Do not touch the sensor head or attenuator cable and connectors while taking
measurements. The probe head is not a hand-held device