Functional check
The functional check verifies the basic operation of the probe functions:
-
Turn on both emitter and receiver.
-
Connect the receiver to a 50Ohm terminated scope and press the OFFSET key.
-
The signal trace should go to zero.
-
Press the ON key on the emitter and check that it turns off.
Connect the XFVP-L in a setup as shown below. Install new alkaline batteries in the transmitter.
Turn on the probe head and receiver and wait 5 minutes before starting the tests.
OFFSET Verification
.
This step can also be carried out with a short at the input of the probe
1) Set the voltage source to 0V.
MEASURES TAKEN BY
Model
Serial #
Calibration date
BENCH MULTIMETER
POWER SUPPLY/
VOLTAGE SOURCE
* Does not need calibration
since measurements are
carried out with the DMM
BR-XFVPL-Rev 2. 12/22