GA FM 1 03.05 • 22
/
11
stoRaGe anD tRanspoRtation e
TraNSPOrT
If the FlowMarker will not be needed for more than 3 months and will not be connected
to the power supply during this time, charge the battery completely and take it out of
the device.
Such preparation makes the battery storable for up to 2 years without suffering any
damage.
During storage and transportation strictly prevent the battery contacts from being
short-circuited by conductive materials such as metal, carbon fiber, graphite, foils etc.
otherwise danger of fire.
For transport a special case is available.
Especially when wrapping or packingthe FlowMarker (e.g. as flight baggage), take the
battery out of the device at any rate to prevent the device from starting.
The two contacts of the battery must also be covered (e. g. insulating tape etc.).
It is unavoidable that some condensation will occur in the fog application tube during
intensive use of the FlowMarker.
In well ventilated warm areas a possible fluid condensation, for example on a could
surface, will dry up without leaving a residue. In the application tube however it is likely
that under intensive use the condensed fluid film will increase and fluid may drop out
the tube.
To avoid a slide risk
condensation absorbing rings are available as an accessory,
which are inserted at both ends of the application tube.
The smaller ring
KR 1 is to be inserted from the muzzle side
of
the black tube.
The larger ring
KR 2 has to be inserted from the „threaded“ side
of the application tube until it fits in
behind the air inlet holes
at
the recess.
The rings will absorb possible fluid condensation and may dry out
if the FlowMaker is not used. However they have to be replaced
periodically depending on th intensity of use, at the latest when they
are completely soaked.
To
remove the “front” ring KR1, use the attached plastic tube
and push it out from the rear end.
To remove the KR2 ring, flush the tube inside with warm water
and let the ring get soft before pulling it out.
If necessary put the applicator tube end in warm water for some 15
min., remove the ring, clean it from fluid residues and other impuri-
ties and wipe the tube completely dry before mounting it back to
the FlowMarker.
CONDENSaTION aB-
SOrBING rINGS
Содержание Tintschl Engineering FlowMarker
Страница 25: ...GA FM 1 03 05 22 25...
Страница 26: ...26 22 GA FM 1 05 01...
Страница 27: ...GA FM 1 03 05 22 27...
Страница 28: ...28 22 GA FM 1 05 01...