Doc no: M18 Issue no: 07 Created: 24/08/2009
9
Good filter seating is essential for optimum performance.
-
When all filters are in place, tighten the clamping nuts firmly.
-
One filter identification label will be supplied with the filter so that when the filter is
installed it is visible through the viewing window.
-
Replace both the front cover panels and retaining screws, ensuring the lighting plug is
reconnected.
-
Access to the pre-filters is made when the internal upper baffle is unfastened at the
front and allowed to fold down. Access does not interrupt the seating of the main
filters.
-
Carefully unscrew the knurled nuts holding the internal baffle in position and allow it
to fold down to expose the pre-filter carrying frames.
-
Unfasten the knurled nut holding the lower pre-filter carriers and allow them to fold
down.
-
Beneath each main filter place a single FILTRETE particle pre-filter.
-
Fold shut the pre-filter carriers and then the internal baffle.
-
Amend the Operational Safety Log Book.
-
Connect to the power supply and the AIRONE Filtration Fume Cupboard is ready for
use.