Doc no: M18 Issue no: 07 Created: 24/08/2009
11
OPERATING INSTRUCTIONS.
-
Check voltage on data plate.
-
Insert mains plug into power supply socket.
-
Check filters fitted before operating.
-
Turn the mains switch to the ‘ON’ position. Press “fan on” button on PCB. The fan
motor(s) will run.
-
The AIRONE Filtration Fume Cupboard should be run for a test period of one hour with
the sash in the lowest position.
-
Do not use physically or chemically incompatible substances within the working area or
fume enclosure.
-
Follow the filter instructions and recommendations as detailed in the Operational
Safety Log Book.
-
If the correct filters are not available - do not operate the AIRONE Filtration Fume
Cupboard in the recirculatory mode.
-
The remaining filter efficiency should be regularly checked as detailed in the
Operational Safety Log Book. Filters should be replaced before the maximum
breakthrough value in the Filter Test procedure is reached.
-
Spare filters should be kept in stock to ensure that filters can be replaced in good time
during regular monitoring or for emergency installations - should a spillage of volatile
materials occur.
-
Regular and careful cleaning of the work surface, back baffle and support channels is
recommended to prevent a build up of contaminant substances.
Following these instructions will ensure protection for the operator, environment, and also
prolong the life of the filters.
It is the responsibility of the operator to ensure correct use of the AIRONE Filtration Fume
Cupboard at all times.
Содержание AIRONE 1000R
Страница 5: ...Doc no M18 Issue no 07 Created 24 08 2009 5 Airone R airlfow ...
Страница 19: ...Doc no M18 Issue no 07 Created 24 08 2009 19 Appendix VI Main Block Wiring Diagram ...
Страница 20: ...Doc no M18 Issue no 07 Created 24 08 2009 20 Appendix VII Control System ...
Страница 22: ...Doc no M18 Issue no 07 Created 24 08 2009 22 Appendix VIII Siting Locations for Airone R Range Figure 16 ...
Страница 23: ......