PRINCIPLE OF OPERATION
(1) Recirculatory Mode of Operation (see diagram on following page)
Air is drawn in through the front opening (1) at a velocity ensuring containment of any
contaminants generated within the work area (2).
The operator can determine the required aperture by sliding the glass sash up to a maximum
determined by the sash stop.
The contaminants from area (2) are drawn either upwards as in the case of lighter than air
volatiles (4), or drawn across the worksurface to the secondary extract slot created by the
back baffle, as in the case of heavier than air volatiles or particulates (3).
The combinations of extract slots (6) and (7) ensure that all contaminants entrained in the
airflow are effectively transferred to the filter (5).
The filter(s) (5) combine large surface areas with a deep bed depth to ensure efficient
filtration of contaminants. Clean filtered air is then drawn through the fan(s) (8) and after
passing into the dispersal chamber (9) is exhausted via the louvres (10).
The relatively high pressure drop across the carbon filters ensures even flow through the
filter and exceptionally uniform saturation.
AIRONE™ R Filtration Fume Cupboards present a combination of advantages and unique
features - ensuring the health and safety of operators by prevention of their exposure to
concentrations of toxic chemical compounds providing protection to their environment - in
line with British, European and US Federal Safety Standards.
Moving the vertical gliding safety glass upwards and downwards causes the fan to speed up
or slow down to give a constant face velocity. To avoid the unit alarming unnecessarily, the
sash should be moved slowly.
(2) Ducted - Vented Mode of Operation:
Should venting to outside atmosphere be required after removal of contaminants by efficient
filtration, the duct-work can be connected to the AIRONE Filtration Fume Cupboard via an
optional duct adapter top box fitted with a 250 mm dia. duct outlet and internal transition
adaptors fitted onto the fan(s). This Ducted Filtration Fume Cupboard will also protect
personnel and the environment outside the building - contributing to the overall reduction of
atmospheric pollution.
Содержание AIRONE 1000R
Страница 5: ...Doc no M18 Issue no 07 Created 24 08 2009 5 Airone R airlfow ...
Страница 19: ...Doc no M18 Issue no 07 Created 24 08 2009 19 Appendix VI Main Block Wiring Diagram ...
Страница 20: ...Doc no M18 Issue no 07 Created 24 08 2009 20 Appendix VII Control System ...
Страница 22: ...Doc no M18 Issue no 07 Created 24 08 2009 22 Appendix VIII Siting Locations for Airone R Range Figure 16 ...
Страница 23: ......