Airone R Range Examination Record
Site:.....................................................................
..............................................................…………
Air flow meter used:.........................................……
…………………………………………………………...
Date
of
Check
Readings in each of the cells as
above.
Average
( m/s )
Pre Filters
Clean?
Any deterioration or
damage observed.
Fail if face velocity
Initials
of
Tester
a b c d e f g h i
Yes/No
< 0 . 3 m / s
If a drop of 10% is found then the cause of the problem should be cleared. (Check Pre-filters)
If the minimum face velocity is below 0.3 m / s then airflow in the fume cupboard is inadequate.