26
1.
Make sure the oxygen supply is OFF and vaporizer is set to zero.
2.
Make sure the anesthesia system is in a well-ventilated area.
3.
Attach a drain tube to the drain port. Place the other end of the tube in a receptacle
to catch the drained anesthetic agent.
4.
Loosen the sealing cap, then loosen the bottom screw with a screwdriver to make
the anesthetic agent flows naturally.
1)
When the anesthetic agent stops draining from the vaporizer, tighten the bottom
screw and sealing cap.
2)
Remove the drain tube and deal with the waste anesthetic agent according to the
regulations.
Do not wipe spilled anesthetic agent from any painted surfaces or the
finish may be damaged. Allow the anesthetic to evaporate.
Perform this procedure in a well-ventilated location.
Refer to the anesthetic agent manufacturer’s MSDS for required
personal protective equipment and handling and disposal of waste
anesthetic agent.
Do not mix the anesthetic agent with other liquid.