16
4.3.2.2 Easy Fill
1)
Turn and remove the sealing cap of
the vaporizer. Check the black
sealing ring.
2)
Mount an adaptor onto the isoflurane
bottle, insert into filler port along the
groove, and press it in the end.
3)
Observe the agent level sight glass
on the vaporizer. Keep the agent
level inside the tube between the
two marks.
4)
Tighten the sealing cap.