operator can confirm the respiratory status with the floatable discs. 

4  

Endotracheal tube/ mask connector plug 

Used for placing the endotracheal tube/ mask connector. Meanwhile the plug blocks the 

breathing circuit, it’s useful for system leak detection, adjusting system pressure and 

flushing circuit. 

5

  CO

2

 absorbent canister 

The absorbent canister is filled with a granular carbon dioxide absorbent material eg. soda 

lime. It uses a coaxial gas flow path to provide for more efficient use of absorbent, while 

reducing flow resistance. The chemical reaction that takes place in the canister creates heat 

and humidity, which is added to the delivery of anesthesia to the animal. 

6

  Vaporizer 

The  vaporizer is the most  complicated and expensive part of the anesthesia system. More 

attention  should  be  paid  to  its  maintenance  and  operation.  It  converts  a  liquid  anesthetic 

agent into a vapor which is added to oxygen. The anesthetic vapor is measured in volume 

percent (vol %). A dial on top of the vaporizer allows the operator to select the amount of 

vapor administered to  the animal.  The  coiled pipe design inside the vaporizer ensures the 

output  is  not  affected  by  the  fluctuation  of  pressure,  temperature  and  gas  flow.  The  inner 

design of the vaporizer is different according to the anesthetic agent. Please use the correct 

anesthetic agent  following  the tag on the vaporizer. The misuse  may cause the damage to 

the equipment and animal. 

7

  Pressure gauge 

Measures and displays the pressure of gas mixture in breathing circuit. 

8

  O

2

 flush button 

Press  to  deliver  the  oxygen  directly  to  breathing  circuit.  It  is  generally  used  to  flush  the 

anesthetic gas in system or in emergency. 

9

  Rebreathing circuit /Non-rebreathing circuit tube socket 

Used for connecting to a non-rebreathing circuit, which is generally applicable to animals 

below 7 kg. 

 

10  

Relief valve 

Operator can determine the inspiratory pressure for ventilating the animal by adjusting the 

valve  while  observing  the  system’s  pressure  gauge.  When  the  pressure  of  gas  mixture 

exceeds  the  setting,  the  gas  will  be  expired  through  valve  to  keep  the  pressure  stable. 

Содержание R640-S1

Страница 1: ...R640 S1 Flexible User Manual V1 1 Veterinary Anesthesia Machine...

Страница 2: ...emark or mark in this instruction manual are only for editing purposes without other purposes the rights belong to their respective rights owners Statement RWD reserves the right to modify the content...

Страница 3: ...h 7 2 5 System malfunction 7 3 Unpacking and assembling 8 3 1 Overview 8 3 2 Unpacking the system 8 3 3 Assembling the system 8 3 3 1 Assembling base 8 3 3 2 Assembling main support and vaporizer 9 3...

Страница 4: ...eping 20 6 5 Problem Solution matrix 20 7 Maintenance 24 7 1 Overview 24 7 2 Safety 24 7 3 Annual inspection 24 7 4 Machine status 24 7 5 Record keeping 24 7 6 Parts and materials 24 7 7 Cleaning the...

Страница 5: ...or contact us immediately 1 2 Safety Operation of anesthesia system involves oxygen gas pressure and the use of potentially hazardous materials In order to prevent injury to animals or operators and...

Страница 6: ...Selectatec The concentration of anesthetic gas from the vaporizer can be adjusted from 0 to 5 Isoflurane and 0 to 8 Sevoflurane without effect by fluctuation of flux temperature flow rate and pressure...

Страница 7: ...zer 12 Rebreathing circuit connector 1 Main support Solid metal support for assembling of following component 2 O2 flowmeter Control the oxygen flow rate Turn the O2 flowmeter control valve to regulat...

Страница 8: ...percent vol A dial on top of the vaporizer allows the operator to select the amount of vapor administered to the animal The coiled pipe design inside the vaporizer ensures the output is not affected b...

Страница 9: ...nister 11 Breathing bag mount Used for connecting to a breathing bag when using the anesthesia system in a rebreathing circuit configuration 12 Rebreathing circuit joint Used for connecting to a rebre...

Страница 10: ...ection provides requirements necessary to ensure safe and reliable operation of your R640 S1 anesthesia system 2 4 1 Animal and operator safety Always keep a backup anesthesia delivery plan in case of...

Страница 11: ...er only with the anesthetic agent for which it is designed Avoid use of oil or grease on any anesthesia or respiratory equipment where oxygen is used An explosion may occur 2 4 2 Preventing system and...

Страница 12: ...ure that all boxes have been received Examine the shipping boxes for damage Immediately make a damage claim to the carrier if there is serious damage and contact RWD Taking photos is suggested for kee...

Страница 13: ...9 3 3 2 Assembling main support and vaporizer Step1 Step2 Step3 An example of the installing method for Cage Mount type vaporizer is showed as below Step1 Step2 Step3 Step4...

Страница 14: ...method for Selectatec type vaporizer is showed as below Step1 Step2 Step3 Step4 The detail picture is shown below using the Selectatec type as an example the Cage Mount type is also applicable 1 Tube...

Страница 15: ...11 3 3 3 Assembling CO2 absorber canister Push the canister to the end of the track Step1 Step2 3 3 4 Assembling monitor tray and gas filter canister Optional Step1 Step2 Step3...

Страница 16: ...12 3 3 5 Assembling breathing bag rebreathing circuit and non rebreathing circuit Rebreathing circuit Non rebreathing circuit Switching valve Connected to the gas recovery tank...

Страница 17: ...ister with CO2 absorbent Do not exceed the maximum capacity scale Reinstall the absorbent canister and tighten the knob 3 System leak checking Close all the open end of anesthesia system Make sure vol...

Страница 18: ...bt or need any help contact RWD or local dealer immediately 4 3 2 Filling the anesthetic vaporizer User choose different vaporizer the anesthetic filling method is also different RWD supply three kind...

Страница 19: ...the sealing cap of the vaporizer Do not leave any debris 2 Carefully pour appropriate agent into the filler port Observe the agent level sight glass on vaporizer Keep agent level inside tube between t...

Страница 20: ...k the black sealing ring 2 Mount an adaptor onto the isoflurane bottle insert into filler port along the groove and press it in the end 3 Observe the agent level sight glass on the vaporizer Keep the...

Страница 21: ...l lever is tightened and then insert the adapter into the filer port rotate the upper jacking bar clockwise to tighten it and lift the anesthetic bottle up and keep upright to add the agent 3 Upon com...

Страница 22: ...otracheal intubation for animal after anesthesia induction 2 Turn the O2 control valve counterclockwise and observe the position of float to set a suitable flow rate 3 Press the dial lock key and turn...

Страница 23: ...o the needle of pressure gauge Stop rotating until it reaches appropriate pressure which is recommended not to be over 20 cmH2O 5 Turn off the flowmeter and reconnect the breathing bag and rebreathing...

Страница 24: ...uring the procedure basic personal protective is necessary such as wearing gloves mask and eye protection 6 3 Machine status Unless otherwise specified the anesthesia system may be connected to gas su...

Страница 25: ...the dial lock key on the vaporizer and adjust the dial to the desired volume percent of anesthetic d Leak around the vaporizer filler port Make sure the sealing cap on the vaporizer is fully closed e...

Страница 26: ...ary 2 Check proper function of inspiration and expiration valves e Leak around the vaporizer filler port Make sure the sealing cap on the vaporizer is fully closed 6 Animal sleep level is too deep a A...

Страница 27: ...t port is fully closed 2 Verify that inlet port is not obstructed by animal hair etc c Leak in animal breathing circuit Check all hoses connections particularly around mask or endotracheal tube d CO2...

Страница 28: ...1 anesthesia system be inspected annually for proper function Contact RWD or your local dealer for service 7 4 Machine status Unless otherwise specified the anesthesia system may be connected to gas s...

Страница 29: ...Wipe the canister gasket located on the underside of the machine with a clean cloth dampened with germicidal cleaner Make sure there is no absorbent residue remaining on the gasket surface d Fill the...

Страница 30: ...ly 1 When the anesthetic agent stops draining from the vaporizer tighten the bottom screw and sealing cap 2 Remove the drain tube and deal with the waste anesthetic agent according to the regulations...

Страница 31: ...ation range Isoflurane 0 5 5 v v Sevofluran 0 5 8 v v Settable concentration point Isoflurane 0 0 5 1 0 1 5 2 0 2 5 3 0 3 5 4 0 4 5 5 0 v v Sevofluran 0 0 5 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 v v Perfusi...

Страница 32: ...e unit conversions Table 9 1 Pressure unit conversions 9 4 Minimum alveolar concentration MAC anesthetic levels Table 9 2 MAC levels Animal Halothane Isoflurane Sevoflurane Cat 1 19 1 63 2 58 Dog 0 87...

Страница 33: ...Blvd San Diego CA 92121 USA Add 1 2 Floor Robeta Building No 11 High Tech North 4th Road Nanshan District Shenzhen Guangdong 518057 China Tel 001 858 900 6602 86 755 86111286 After sales Service 86 7...

Отзывы: