24
(5) Capacitance
Frequency
Range
Accuracy
100Hz
120Hz
2000.0pF
± (0.5% + 5d)
[2]
20.000nF
± (0.2% + 5d)
200.00nF
2000.0nF
20.000uF
200.00uF
± (0.5% + 5d)
2000.0uF
± (1.0% + 5d)
20.000mF
± (2.0% + 5d)
[2]
1KHz
2000.0pF
± (0.5% + 5d)
[2]
20.000nF
± (0.2% + 5d)
200.00nF
2000.0nF
20.000uF
± (0.5% + 5d)
200.00uF
± (1.0% + 5d)
2000.0uF
± (2.0% + 5d)
[2]
10KHz
200.00pF
± (0.5% + 5d)
[2]
2000.0pF
± (0.2% + 5d)
20.000nF
200.00nF
2000.0nF
± (0.5% + 5d)
20.000uF
± (2.0% + 5d)
200.00uF
± (5.0% + 5d)
[2]
100KHz
[1]
20.000pF
± (0.5% + 20d)
[2]
200.00pF
± (0.2% + 5d)
2000.0pF
20.000nF
200.00nF
± (0.5% + 5d)
2000.0nF
± (2.0% + 5d)
20.000uF
± (5.0% + 5d)
[2]
[1] The 100KHz test frequency is only at LCR1703.
[2] The measuring time is 2 seconds.
Input Protection
:
30V
DC
or 30V
AC
rms
Minimum Resolution
:
0.001pF in the 20.000pF range.
Measuring Time
:
800ms
2-wire Accuracy
:
Add 1.0% to accuracy.
Note : If D > 0.1, the accuracy should be multiplied by
2
1
D
Содержание LCR 1701
Страница 1: ...EN Instruction Manual LCR 1701 LCR 1703 Digital Multimeter...
Страница 7: ...6 Phase Drawing...
Страница 8: ...7 Making 5 wire measuring with the SMD test probe...
Страница 9: ...8 Making 4 wire measuring in the 5 wire terminal Making 2 wire measuring with the alligator clip set...
Страница 11: ...10 Measuring D Q ESR Press the D Q ESR button to select the measuring function...