25
(6) Resistance
Frequency
Range
Accuracy
100Hz
120Hz
200.00
Ω
± (0.2% + 5d)
2.0000K
Ω
20.000K
Ω
200.00K
Ω
2.0000M
Ω
20.000M
Ω
± (0.5% + 5d)
200.00M
Ω
± (1.0% + 5d)
[2]
1KHz
20.000
Ω
± (0.5% + 15d)
[2]
200.00
Ω
± (0.2% + 5d)
2.0000K
Ω
20.000K
Ω
200.00K
Ω
2.0000M
Ω
20.000M
Ω
± (2.0% + 5d)
200.00M
Ω
± (5.0% + 5d)
[2]
10KHz
20.000
Ω
± (0.5% + 15d)
[2]
200.00
Ω
± (0.2% + 5d)
2.0000K
Ω
20.000K
Ω
200.00K
Ω
2.0000M
Ω
± (2.0% + 5d)
20.000M
Ω
± (5.0% + 5d)
100KHz
[1]
20.000
Ω
± (0.5% + 15d)
[2]
200.00
Ω
± (0.2% + 5d)
2.0000K
Ω
20.000K
Ω
200.00K
Ω
± (2.0% + 5d)
2.0000M
Ω
± (5.0% + 5d)
[1] The 100KHz test frequency is only at LCR1703.
[2] The measuring time is 2 seconds.
Input Protection
:
30V
DC
or 30V
AC
rms
Minimum Resolution
:
0.001
Ω
in the 20.000
Ω
range.
Measuring Time
:
800ms
2-wire Accuracy
:
Add 1.0% to accuracy.
Note
:
If Q > 0.1, the accuracy should be multiplied by
2
1
Q