Turn off steam supply to sterilizer and wait until jack-
et and chamber pressure is 0 psi.

Note that there are three connections to be made to
the WATER-MIZER:

• Cold water inlet

• Hot water inlet

• Tempered water outlet to drain

Position the WATER-MIZER to allow the most direct
path of piping to minimize fittings.

Mount the WATER-MIZER by attaching the mounting
plate to the adjustable floor stand (see page 8 to
order the Adjustable Floor Stand, RPI Part
#RPH470), or by attaching the WATER-MIZER
directly to the frame of the sterilization equipment.
See page 4 for mounting options.

DO NOT SUPPORT WATER-MIZER ENTIRELY BY
PIPING. MOUNT WATER-MIZER FIRMLY BEFORE
PIPING.

Position unions on all connections as close to the
WATER-MIZER as possible to make cleaning and
maintenance easier. (For Maintenance Instructions,
see page 6.)

Cold water supply connection instructions:

• Cold water supply connection on valve is 3/8" FPT.

• Disconnect jacket and chamber trap cooling valves

on sterilizer and plug all openings.

• Verify that the supply water pressure to the valve

will be at least 25 psi (172 kPa) and not more than
80 psi (552 kPa).

• Pipe a 3/8" line directly to the WATER-MIZER from

the main water supply line.

• Install a cold water supply union as close to the

WATER-MIZER as possible.

• Install a cold water shut-off valve before the union

in the cold water supply line.

Installation checklist.

Installation

Hot water inlet connection instructions:

• Hot water inlet connection is 1" FPT.

• Using Teflon

®

tape or pipe sealant, install

brass tee and vacuum breaker assembly pro-
vided, to the sterilizer drain inlet on the
WATER-MIZER.

• Locate a union as close to the WATER-

MIZER as possible

• Connect all sterilizer drain lines to 1" pipe

and run pipe as directly as possible to the
WATER-MIZER. If the piping to the sterilizer
drain inlet has a horizontal run, maintain a
pitch to the WATER-MIZER of at least 1/8"/ft
(1%).

STERILIZER DRAIN LINES MAY BE HOT!
USE CAUTION. MAKE SURE STEAM SUP-
PLY IS SHUT OFF.

Tempered water outlet connection instructions:

• Tempered water outlet connection is 2" MPT.

• Install a union as close to the WATER-MIZER

as possible.

• Run a 2" pipe as directly as possible from the

WATER-MIZER to the drain. Maintain a pitch
to drain of at least 1/8"/ft (1%).

• Make sure there is a 1" (25 mm) air gap

between the drain piping and the drain.

Fill the WATER-MIZER tank with cold water
before initial use.

If the WATER-MIZER is run

empty, excessive noise may occur.

Turn on sterilizer and using the built-in
thermometer, monitor drain water temperature.
Adjust valve on case as indicated to maintain
drain temperature below 140°F (60°C).

Installation

:

Installation checklist

3

Step-by-step installation instructions

Содержание WATER-MIZER RPA466

Страница 1: ...d and save these instructions WATER MIZER Water saving tempering device for sterilization equipment Installation operation and maintenance manual Proud to be an ISO 9001 certified company RPI Part RPA...

Страница 2: ...ill have questions call our technical support group at 800 221 9723 x135 Technicians are available from 8 00 a m to 4 30 p m Pacific Time Monday through Friday excluding major holidays For information...

Страница 3: ...s 2 Warranty 9 Installation Installation instructions 3 Mounting 4 WATER MIZER table of contents Table of Contents Operation Start up and operation instructions 5 Maintenance Maintenance instructions...

Страница 4: ...R MIZER through piping connected to the top threaded connection A vacuum breaker prevents backflow into potable water systems Cold water enters through the temperature actuated valve The valve and the...

Страница 5: ...G PLATE TO CENTER OF HOT WATER INLET 9 1 2 241 H HEIGHT FROM CENTER OF HOT WATER INLET TO TOP OF VACUUM BREAKER 3 76 J HEIGHT FROM BOTTOM OF MOUNTING PLATE TO CENTER OF TEMPERED WATER OUTLET 4 1 4 108...

Страница 6: ...o the WATER MIZER as possible Install a cold water shut off valve before the union in the cold water supply line Installation checklist Installation Hot water inlet connection instructions Hot water i...

Страница 7: ...hnique Mounting the WATER MIZER to the frame may require that holes be drilled in the frame Secure WATER MIZER to the frame of the sterilizer by attaching the integral mounting plate to the frame Note...

Страница 8: ...valve s sensor located inside the chamber near the outlet ensures that water leaving the WATER MIZER is less than 140 F 60 C before entering the municipal sewer system Tempered water at 140 F 60 C or...

Страница 9: ...mineral accumulation has occurred remove the split nut holding the sensor in place and remove the thermal sensor from the WATER MIZER chamber Gently clean the sensor with an abrasive pad Do not twist...

Страница 10: ...sor from WATER MIZER and gently remove mineral accumulation with an abrasive pad Do not twist capillary tubing during removal or cleaning Valve is not set properly Adjust valve so drain effluent is be...

Страница 11: ...ve shown below When you order a Replacement Valve with Sensor RPI Part RPV468 you will receive the style shown to the right RPV468 Important Note For Replacement Valve with Sensor please see Important...

Страница 12: ...Any warranty claim must be submitted to RPI in writing within the stated warranty period Two year limited warranty RPI s limited warranty is made in lieu of and RPI dis claims all other warranties wh...

Отзывы: