Troubleshooting guide.

Maintenance

Water leaving the
WATER-MIZER is
hotter than 140°F
(60°C)

Excessive Noise

Thermometer Not
Reading Properly

Mineral accumulation in
mixing chamber

Mineral accumulation on
thermal sensor

Cold water valve setting

Cold water supply
Less than 25 psi

Valve malfunction

Steam trap on sterilizer
stuck open or defective

Dry Water-Mizer

Mineral Accumalation on
thermometer

Remove the WATER-MIZER and
dislodge mineral accumulation.

Remove thermal sensor from WATER-
MIZER and gently remove mineral
accumulation with an abrasive pad.

Do not twist capillary tubing during
removal or cleaning.

Valve is not set properly. Adjust valve so
drain effluent is below 140°F.

Turn on cold water supply and make
sure cold water is present.

Replace valve or use PM Kit.

Have sterilizer serviced by a qualified
technician.

Water-Mizer must be filled with cold
water before initial use or before
returning sterilizer to service.

Continued

operation will correct this problem.

Remove thermometer from WATER-
MIZER and gently remove mineral
accumulation with an abrasive pad.

Problem

Number

Problem Cause

Action

Problem

Maintenance:

Troubleshooting guide

7

Содержание WATER-MIZER RPA466

Страница 1: ...d and save these instructions WATER MIZER Water saving tempering device for sterilization equipment Installation operation and maintenance manual Proud to be an ISO 9001 certified company RPI Part RPA...

Страница 2: ...ill have questions call our technical support group at 800 221 9723 x135 Technicians are available from 8 00 a m to 4 30 p m Pacific Time Monday through Friday excluding major holidays For information...

Страница 3: ...s 2 Warranty 9 Installation Installation instructions 3 Mounting 4 WATER MIZER table of contents Table of Contents Operation Start up and operation instructions 5 Maintenance Maintenance instructions...

Страница 4: ...R MIZER through piping connected to the top threaded connection A vacuum breaker prevents backflow into potable water systems Cold water enters through the temperature actuated valve The valve and the...

Страница 5: ...G PLATE TO CENTER OF HOT WATER INLET 9 1 2 241 H HEIGHT FROM CENTER OF HOT WATER INLET TO TOP OF VACUUM BREAKER 3 76 J HEIGHT FROM BOTTOM OF MOUNTING PLATE TO CENTER OF TEMPERED WATER OUTLET 4 1 4 108...

Страница 6: ...o the WATER MIZER as possible Install a cold water shut off valve before the union in the cold water supply line Installation checklist Installation Hot water inlet connection instructions Hot water i...

Страница 7: ...hnique Mounting the WATER MIZER to the frame may require that holes be drilled in the frame Secure WATER MIZER to the frame of the sterilizer by attaching the integral mounting plate to the frame Note...

Страница 8: ...valve s sensor located inside the chamber near the outlet ensures that water leaving the WATER MIZER is less than 140 F 60 C before entering the municipal sewer system Tempered water at 140 F 60 C or...

Страница 9: ...mineral accumulation has occurred remove the split nut holding the sensor in place and remove the thermal sensor from the WATER MIZER chamber Gently clean the sensor with an abrasive pad Do not twist...

Страница 10: ...sor from WATER MIZER and gently remove mineral accumulation with an abrasive pad Do not twist capillary tubing during removal or cleaning Valve is not set properly Adjust valve so drain effluent is be...

Страница 11: ...ve shown below When you order a Replacement Valve with Sensor RPI Part RPV468 you will receive the style shown to the right RPV468 Important Note For Replacement Valve with Sensor please see Important...

Страница 12: ...Any warranty claim must be submitted to RPI in writing within the stated warranty period Two year limited warranty RPI s limited warranty is made in lieu of and RPI dis claims all other warranties wh...

Отзывы: