![RPI WATER-MIZER RPA466 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/rpi/water-mizer-rpa466/water-mizer-rpa466_installation-operation-and-maintenance-manual_1495540010.webp)
1
2
3
Troubleshooting guide.
Maintenance
Water leaving the
WATER-MIZER is
hotter than 140°F
(60°C)
Excessive Noise
Thermometer Not
Reading Properly
Mineral accumulation in
mixing chamber
Mineral accumulation on
thermal sensor
Cold water valve setting
Cold water supply
Less than 25 psi
Valve malfunction
Steam trap on sterilizer
stuck open or defective
Dry Water-Mizer
Mineral Accumalation on
thermometer
Remove the WATER-MIZER and
dislodge mineral accumulation.
Remove thermal sensor from WATER-
MIZER and gently remove mineral
accumulation with an abrasive pad.
Do not twist capillary tubing during
removal or cleaning.
Valve is not set properly. Adjust valve so
drain effluent is below 140°F.
Turn on cold water supply and make
sure cold water is present.
Replace valve or use PM Kit.
Have sterilizer serviced by a qualified
technician.
Water-Mizer must be filled with cold
water before initial use or before
returning sterilizer to service.
Continued
operation will correct this problem.
Remove thermometer from WATER-
MIZER and gently remove mineral
accumulation with an abrasive pad.
Problem
Number
Problem Cause
Action
Problem
Maintenance:
Troubleshooting guide
7