Inner conductor protrusion 

 

This  may  result  in  buckling  of  the  socket  contacts  or  damage  to  the  internal  structure  of  a 
device due to the axial forces generated. 

 

NOTE: 

At no time should the pin depth of the connector be protruding

 
 

Inner conductor recession 

 

This will result in poor reflection and possibly unreliable contact. 

 

6.2.2 Gauges 

 
The  Rosenberger  gauges  are  of  screw-on  type  in  plug  and  jack  version.  They  can  be  used  to 
gauge calibration standards as well as test ports, test port cables, adapters and DUTs.  
 
Naming convention:  
A “gauge plug” (male) is used to measure a DUT with plug (male) connector. The gauge itself has 
a jack (female) connector and needs a gauge block with plug (male) connector. For a “gauge jack” 
(female) all gender are vice versa. 
 
Checking devices for protrusion or excessive recession only a single measurement is sufficient. 
 
More  precise  measurements  or  checking  devices  very  close  to  critical  limits  should  be  done  in 
different orientations averaging up to 4 single measurements. 
 
When  a  gauge  block  is  used  to  initially  adjust  the  gauge  a  torque  spanner  should  be  used  to 
tighten up the connection to the correct torque. 
 

6.2.3 Connector gauge measurement resolution 

 

Because of connector gauge measurement resolution uncertainties (one small division on the dial) 
and variations in measurement technique from user to user connector dimensions may be difficult 
to measure. Dirt and contamination can cause differences of 0.005 mm and, in addition, the way 
that the gauge is used can result in larger variations. When using a gauge system for mechanical 
compliance testing of connectors carry out the following procedure each time: 
 

  Carefully inspect the connector to be tested and clean it if necessary. 

  Clean and inspect the gauge and the gauge block. 

  Carefully zero the gauge with the gauge block in place. 

  Remove the gauge block. 

 

  Measure the connector using the gauge and note the reading. 

  Repeat the process at least once or more times as necessary. 

 
 

Содержание 02CK001-150

Страница 1: ...mbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 17 Dieses Dokument ist urheberrechtlich gesch tzt This document is protected by...

Страница 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installa...

Страница 3: ...ed Contributing factors are noise linearity and temperature stability of the VNA transmission and reflection stability of a test port cable as well as operator experience and care VNAs usually offer s...

Страница 4: ...C10S3 1 Calibration load jack 02K150 C10S3 1 Calibration adaptor plug plug 02S121 S20S3 1 Calibration adaptor jack jack 02K121 K20S3 1 Combi wrench 03W008 000 5 mm 6 mm 7 mm 8 mm 9 mm 1 Torque wrench...

Страница 5: ...recautions and information for installing Standard Definitions on a Vector Network Analyzer Calibration Certificate See chapter Calibration Certificate User Manual This document 1 5 Calibration Certif...

Страница 6: ...ns are defined for all calibration standards gauges and the torque wrench in the specific data sheets 3 Standard Definitions 3 1 Standard Definitions Standard definitions provide the data needed to ma...

Страница 7: ...a more complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously...

Страница 8: ...setting parameters Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the d...

Страница 9: ...vailable In cases when the two VNA test ports have different connector systems installed e g RPC 3 50 and RPC 2 40 UOSM can handle this situation A second calibration kit is needed Calibration Procedu...

Страница 10: ...calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following services are...

Страница 11: ...ood connector repeatability and its effect on the overall uncertainty of a measurement procedure should encourage careful inspection interface gauging and handling of coaxial connectors Prior to use a...

Страница 12: ...wear at least a grounded wrist strap and observe correct procedures 6 2 Interface Dimensions and Gauging It is of the utmost importance that connectors do not damage the test equipment interfaces to...

Страница 13: ...or a gauge jack female all gender are vice versa Checking devices for protrusion or excessive recession only a single measurement is sufficient More precise measurements or checking devices very close...

Страница 14: ...es Dokument ist urheberrechtlich gesch tzt This document is protected by copyright Rosenberger Hochfrequenztechnik GmbH Co KG 6 2 4 Gauge blocks Every connector gauge requires a gauge block that is us...

Страница 15: ...in which the centre conductor extends beyond the outer conductor mating plane This may result in buckling of the socket contacts or damage to the internal structure of a device due to the axial force...

Страница 16: ...l centre conductors are within the observed pin depth values in Table 2 2 Refer to 4 2 Interface Dimensions and Gaging 5 Carefully align the connectors The male connector centre pin must slip concentr...

Страница 17: ...en disconnecting a sliding load leave the centre conductor pullback handle in the locked position Use the combi wrench to prevent the device body from turning Use torque wrench to loosen the connectin...

Отзывы: