4. VNA Calibration 

4.1 VNA Calibration 

 

A  VNA  is  only  as  useful  as  the  accuracy  of  the  measurements  it  makes,  and  this  requires  the 
instrument  to  be  calibrated.  The  calibration  process  employs  a  technique  called  vector  error 
correction,  in  which  error  terms  are  characterized  using  known  standards  so  that  errors  can  be 
removed from actual measurements. The process of removing these errors requires the errors and 
measured quantities to be measured vectorially (thus the need for a vector network analyzer). 
 
 

Main setting parameters

 
It is good practice to start your measurement by selecting the main setting parameters. 
 

  Sweep type (linear / logarithmic) 

  Sweep range (start frequency / stop frequency) 

  Number of points (frequency step size) 

  Source power 

  IF bandwidth (small IF bandwidth basically increases the dynamic range) 

  Sweep time 

 
Connect test port cables and adapters to your VNA as needed for your specific measurement task. 
 
 

Calibration techniques 

 
A  very  large  number  of  calibration  techniques  have  been  developed  to  calibrate  a  VNA.  The 
following list describes calibration techniques you can apply with this calibration kit along with their 
features and applicability for use in specific situations.  
 

One-port calibration 

 

OSL (OSM)

   

 

Open-Short-Load (Open-Short-Match) 

 
When  only  reflection  measurements  are  needed,  the  fastest  and  easiest  calibration 
technique is OSL. It requires 3 connections of calibration standards on the reference plane. 

 

Two-port calibrations 

 

OSLT (TOSM)

 

 

Open-Short-Load-Thru (Thru-Open-Short-Match) 

 
OSLT  is  the  most  commonly  used  two-port  calibration  method.  The  Thru  connections  are 
performed by using well-known, precision calibration adapters in male-male or female-female 
configuration or as a direct connection of the male and the female test ports.  
 
 
 

Содержание 02CK001-150

Страница 1: ...mbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 17 Dieses Dokument ist urheberrechtlich gesch tzt This document is protected by...

Страница 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installa...

Страница 3: ...ed Contributing factors are noise linearity and temperature stability of the VNA transmission and reflection stability of a test port cable as well as operator experience and care VNAs usually offer s...

Страница 4: ...C10S3 1 Calibration load jack 02K150 C10S3 1 Calibration adaptor plug plug 02S121 S20S3 1 Calibration adaptor jack jack 02K121 K20S3 1 Combi wrench 03W008 000 5 mm 6 mm 7 mm 8 mm 9 mm 1 Torque wrench...

Страница 5: ...recautions and information for installing Standard Definitions on a Vector Network Analyzer Calibration Certificate See chapter Calibration Certificate User Manual This document 1 5 Calibration Certif...

Страница 6: ...ns are defined for all calibration standards gauges and the torque wrench in the specific data sheets 3 Standard Definitions 3 1 Standard Definitions Standard definitions provide the data needed to ma...

Страница 7: ...a more complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously...

Страница 8: ...setting parameters Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the d...

Страница 9: ...vailable In cases when the two VNA test ports have different connector systems installed e g RPC 3 50 and RPC 2 40 UOSM can handle this situation A second calibration kit is needed Calibration Procedu...

Страница 10: ...calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following services are...

Страница 11: ...ood connector repeatability and its effect on the overall uncertainty of a measurement procedure should encourage careful inspection interface gauging and handling of coaxial connectors Prior to use a...

Страница 12: ...wear at least a grounded wrist strap and observe correct procedures 6 2 Interface Dimensions and Gauging It is of the utmost importance that connectors do not damage the test equipment interfaces to...

Страница 13: ...or a gauge jack female all gender are vice versa Checking devices for protrusion or excessive recession only a single measurement is sufficient More precise measurements or checking devices very close...

Страница 14: ...es Dokument ist urheberrechtlich gesch tzt This document is protected by copyright Rosenberger Hochfrequenztechnik GmbH Co KG 6 2 4 Gauge blocks Every connector gauge requires a gauge block that is us...

Страница 15: ...in which the centre conductor extends beyond the outer conductor mating plane This may result in buckling of the socket contacts or damage to the internal structure of a device due to the axial force...

Страница 16: ...l centre conductors are within the observed pin depth values in Table 2 2 Refer to 4 2 Interface Dimensions and Gaging 5 Carefully align the connectors The male connector centre pin must slip concentr...

Страница 17: ...en disconnecting a sliding load leave the centre conductor pullback handle in the locked position Use the combi wrench to prevent the device body from turning Use torque wrench to loosen the connectin...

Отзывы: