Reference Manual
00809-0100-4664, Rev BA
January 2010
Rosemount 8712
4-28
Totalizer Units
Totalizer units allow for the configuration of the units that the totalized value
will be displayed as. These units are independent of the flow units.
Measured Gross Total
Measured gross total provides the output reading of the totalizer. This value is
the amount of process fluid that has passed through the flowmeter since the
totalizer was last reset.
To reset the gross total value, you must change the line size. See “Line Size”
on page 3-10 for details on how to change the line size.
Measured Net Total
Measured net total provides the output reading of the totalizer. This value is
the amount of process fluid that has passed through the flowmeter since the
totalizer was last reset. When reverse flow is enabled, the net total represents
the difference between the total flow in the forward direction less the total flow
in the reverse direction.
Measured Reverse Total
Measured reverse total provides the output reading of the totalizer. This value
is the amount of process fluid that has passed through the flowmeter in the
reverse direction since the totalizer was last reset. This value is only totalized
when reverse flow is enabled.
Start Totalizer
Start totalizer starts the totalizer counting from its current value.
Stop Totalizer
Stop totalizer interrupts the totalizer count until it is restarted again. This
feature is often used during pipe cleaning or other maintenance operations.
Reset Totalizer
Reset totalizer resets the net totalizer value to zero. The totalizer must be
stopped before resetting.
NOTE
The totalizer value is saved in the Non-Volatile memory of the electronics
every three seconds. Should power to the transmitter be interrupted, the
totalizer value will start at the last saved value when power is re-applied.
Field Comm.
1, 4, 2, 5, 1
LOI Key
AUX. FUNCTION
Field Comm.
1, 4, 2, 5, 2
LOI Key
TOTALIZE
Field Comm.
1, 4, 2, 5, 3
LOI Key
TOTALIZE
Field Comm.
1, 4, 2, 5, 4
LOI Key
TOTALIZE
Field Comm.
1, 4, 2, 5, 5
LOI Key
START/STOP
Field Comm.
1, 4, 2, 5, 6
LOI Key
START/STOP
Field Comm.
1, 4, 2, 5, 7
LOI Key
READ/RESET
Содержание 8712
Страница 2: ......
Страница 4: ......
Страница 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Страница 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Страница 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Страница 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Страница 131: ...Reference Manual 00809 0100 4664 Rev BA January 2010 B 7 Rosemount 8712 Figure B 1 CSA Installation ...
Страница 132: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 B 8 Figure B 2 Factory Mutual Hazardous Locations ...
Страница 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Страница 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Страница 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Страница 211: ......