ABB SENSORS
Connect coil drive and electrode cables as shown in Figure E-5.
ABB Magmaster MFE
and MFF Sensors (Old
Version) to Rosemount
8712 Transmitter
Figure E-5. Wiring Diagram for
ABB Magmaster MFE and MFF
Sensors (Old Version) and
Rosemount 8712
Table E-3. ABB Magmaster MFE and MFF Sensors (Old Version) Wiring
Connections
Rosemount 8712
ABB Magmaster MFE and
MFF Sensors (Old Version)
1
CD1
2
CD2
17
18
SIG2
19
SIG1
CD1
CD2
SIG GND
DS1
SIG1
SIG2
DS2
ABB MAGMASTER MFE
AND MFF SENSORS
(OLD VERSION)
ROSEMOUNT 8712
TRANSMITTER
This is a pulsed dc magnetic flowmeter.
Do
not connect ac power
to the sensor or
to terminals 1 and 2 of the transmitter
,
or replacement of the electronics board will
be necessary.
Содержание 8712
Страница 2: ......
Страница 4: ......
Страница 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Страница 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Страница 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Страница 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Страница 131: ...Reference Manual 00809 0100 4664 Rev BA January 2010 B 7 Rosemount 8712 Figure B 1 CSA Installation ...
Страница 132: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 B 8 Figure B 2 Factory Mutual Hazardous Locations ...
Страница 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Страница 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Страница 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Страница 211: ......