Reference Manual
00809-0100-4664, Rev BA
January 2010
E-23
Rosemount 8712
Series 1800 (Version 2)
Sensor to
Rosemount 8712
Transmitter
Connect coil drive and electrode cables as shown in Figure E-21.
Figure E-21. Wiring Diagram for
Foxboro Series 1800 (Version 2)
and Rosemount 8712
Table E-19. Foxboro 1800 Version 2 Sensor Wiring Connections
Rosemount 8712
Foxboro
Series 1800 Version 2 Sensors
1
L1
2
L2
Chassis Ground
17
Any Shield
18
Black
19
White
FOXBORO SERIES 1800
SENSOR (VERSION 2)
ROSEMOUNT 8712
TRANSMITTER
White
Black
Shield
Electrode
Connections
GND
L2
L1
Coil Connections
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.
Содержание 8712
Страница 2: ......
Страница 4: ......
Страница 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Страница 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Страница 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Страница 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Страница 131: ...Reference Manual 00809 0100 4664 Rev BA January 2010 B 7 Rosemount 8712 Figure B 1 CSA Installation ...
Страница 132: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 B 8 Figure B 2 Factory Mutual Hazardous Locations ...
Страница 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Страница 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Страница 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Страница 211: ......