Instruction Manual
Maintenance
LIQ-MAN-396R-396RVP
March 2017
8.
Connect the sensor wires to the terminals on the printed circuit board in the junction box
in the manner recommended on the junction box cover,and reattach the BNC connector
to the preamp. Screw on the cover of the junction box aside. Discard sensor tube.
9.
Insert the sensor in the process fitting. Stop it against the closed ball valve. Slide the
processend male connector down the sensor tube to mate with the process fitting. Tighten
the male connector into the process fitting.
10.
Pull back hard on the sensor assembly, as if trying to remove the sensor, to be certain that
the sensor cannot come free from the valve assembly and male connector. The built-in
retraction stop collar at the end of the sensor will butt against the shoulder of the male
connector.
11.
Open ball valve and position the sensor at the desired insertion depth and orientation.
Using a crescent or open end wrench, tighten the hex nut (B) to secure the sensor in place.
See Figure 6-2.
If the sensor is to be stored, the rubber boot should be filled with 7pH buffer solution and replaced
on sensor tip until ready to use.
A stainless steel ferrule is available if the Teflon ferrule does not adequately grip, be careful and avoid over
tightening. This can damage the sensor tube.
NOTICE
If the male connector leaks during insertion or retraction, replace the O-Ring (PN 9550099) in the male
connector body (A).
CAUTION
Maintenance
49
Содержание 396R
Страница 1: ...Rosemount 396R 396RVP pH ORP Sensors Instruction Manual LIQ MAN 396R 396RVP Rev K March 2017 ...
Страница 2: ...hasgkas ...
Страница 6: ...Table of Contents Instruction Manual March 2017 LIQ MAN 396R 396RVP ii Table of Contents ...
Страница 22: ...16 Installation Installation Instruction Manual March 2017 LIQ MAN 396R 396RVP ...
Страница 32: ...Wiring Rosemount 396R Instruction Manual March 2017 LIQ MAN 396R 396RVP 26 Wiring Rosemount 396R ...
Страница 48: ...Wiring Rosemount 396RVP Instruction Manual March 2017 LIQ MAN 396R 396RVP 42 Wiring Rosemount 396RVP ...
Страница 60: ...54 Accessories Diagnostics and Troubleshooting Instruction Manual March 2017 LIQ MAN 396R 396RVP ...