Installation
9
Instruction Manual
Installation
LIQ-MAN-396R-396RVP
March 2017
3.
Thread the male connector body tightly into the ball valve assembly. DO NOT tighten the
hex nut on the male connector body; doing so would not allow the sensor to be inserted
through the ball valve.
4.
Pull back hard on the sensor assembly, as if trying to remove the sensor, to be certain that
the sensor cannot come free of the ball valve assembly. The built-in retraction stop will butt
against the shoulder of the male connector if properly installed.
5.
After confirming that the sensor assembly is properly secured by the valve assembly, the
valve may be opened and the sensor positioned into the process at the desired depth and
orientation.
6.
While holding the sensor in position, tighten the hex nut of the male connector to firmly
secure the sensor in place. When the hex nut is tightened, the Teflon ferrule inside the
compression fitting clamps the sensor tube.
2.3.2
Installation Without Ball Valve
The Rosemount 396R Sensor may be installed through a weldalet or pipe tee or “Y” when used with
a process connector (PN 23166-00 or 23166-01). The sensor should be installed within 80° of
vertical, with the electrode facing down.
The sensor must be captured by the valve assembly and the male connector so that it cannot be blown
free by process pressure if mishandled during insertion or retraction.
CAUTION
Over tightening the hex nut may damage the ferrule.
CAUTION
A stainless steel ferrule is available if the Teflon ferrule does not inadequately grip. When using the
metallic ferrule, care must be taken to avoid over tightening and damaging the sensor tube. If the male
connector leaks during insertion or retraction, replace the O-ring in the male connector.
NOTICE
Figure 2-1: Exploded View of Ball Valve Kit PN 23240-00 used with process connector
PN 23166-00 (or PN 23166-01)
Содержание 396R
Страница 1: ...Rosemount 396R 396RVP pH ORP Sensors Instruction Manual LIQ MAN 396R 396RVP Rev K March 2017 ...
Страница 2: ...hasgkas ...
Страница 6: ...Table of Contents Instruction Manual March 2017 LIQ MAN 396R 396RVP ii Table of Contents ...
Страница 22: ...16 Installation Installation Instruction Manual March 2017 LIQ MAN 396R 396RVP ...
Страница 32: ...Wiring Rosemount 396R Instruction Manual March 2017 LIQ MAN 396R 396RVP 26 Wiring Rosemount 396R ...
Страница 48: ...Wiring Rosemount 396RVP Instruction Manual March 2017 LIQ MAN 396R 396RVP 42 Wiring Rosemount 396RVP ...
Страница 60: ...54 Accessories Diagnostics and Troubleshooting Instruction Manual March 2017 LIQ MAN 396R 396RVP ...