3.1
Wiring for Rosemount 396R
Make electrical connections as shown on Figures 3-1 through 3-15 using the following guidelines.
For wiring Rosemount 396RVP, see Section 4.
1.
Pay particular attention to the analyzer or transmitter model number when following
details on the wiring diagrams to ensure that the connections are made to the proper
terminals.
2.
Use Rosemount custom cable Part Number 9200273 for interconnect.
3.
The maximum distance from the sensor to the transmitter is 15 ft without an integral
preamplifier.
4.
Signal cable should be run in a dedicated conduit and should be kept away from AC power
lines.
Section 3: Wiring Rosemount 396R
For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of
the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the
extension cable to the instrument must be terminated at earth ground or by using an appropriate metal
cable gland fitting, that provides a secure connection to the instrument cable.
NOTICE
Figure 3-1: Cable Preparation Instructions (PN 9200274)
Содержание 396R
Страница 1: ...Rosemount 396R 396RVP pH ORP Sensors Instruction Manual LIQ MAN 396R 396RVP Rev K March 2017 ...
Страница 2: ...hasgkas ...
Страница 6: ...Table of Contents Instruction Manual March 2017 LIQ MAN 396R 396RVP ii Table of Contents ...
Страница 22: ...16 Installation Installation Instruction Manual March 2017 LIQ MAN 396R 396RVP ...
Страница 32: ...Wiring Rosemount 396R Instruction Manual March 2017 LIQ MAN 396R 396RVP 26 Wiring Rosemount 396R ...
Страница 48: ...Wiring Rosemount 396RVP Instruction Manual March 2017 LIQ MAN 396R 396RVP 42 Wiring Rosemount 396RVP ...
Страница 60: ...54 Accessories Diagnostics and Troubleshooting Instruction Manual March 2017 LIQ MAN 396R 396RVP ...