Remote control commands
R&S
®
RTB2000
448
User Manual 1333.1611.02 ─ 11
MMEMory:DELete
<FileSource>
Removes a file from the specified directory.
Setting parameters:
<FileSource>
String parameter
File name and path of the file to be removed. If the path is omit-
ted, the specified file will be deleted in the current directory. Fil-
ters are not allowed.
Example:
Chapter 15.2.1.2, "Saving, copying, and loading setup data"
Usage:
Setting only
MMEMory:DATA
<FileName>,<Data>
Writes data to the specified file in the current directory
reads the data.
Parameters:
<Data>
488.2 block data
The block begins with character '#'. The next digit is the length of
the length information, followed by this given number of digits
providing the number of bytes in the binary data attached.
Parameters for setting and query:
<FileName>
String parameter containing the file name
Example:
MMEM:DATA "abc.txt", #216This is the file
#2: the length infomation has two digits
16: the binary data has 16 bytes.
MMEM:DATA? "abc.txt"
received: #216This is the file
Example:
Chapter 15.2.1.2, "Saving, copying, and loading setup data"
MMEMory:STORe:STATe
<StateNumber>,<FileName>
Saves the current device settings to the specified file in the current directory.
Setting parameters:
<StateNumber>
Range:
1 to 1
Increment: 0
*RST:
1
<FileName>
String parameter
File name, with or without file extension
Example:
MMEM:CDIR "/USB_FRONT/DATA"'
MMEM:STOR:STAT 1,"MORNING.SET"