Remote control commands
R&S
®
RTB2000
338
User Manual 1333.1611.02 ─ 11
ALL
All fields, triggers on the frame start (progressive scanning) or
any field start (interlaced scanning).
LINE
Triggers on the beginning of a specified line in any field. The line
number is set with
ALINe
Triggers on the beginning of all video signal lines.
*RST:
ALL
Manual operation:
See
TRIGger:A:TV:LINE
<Line>
is set to LINE.
Parameters:
<Line>
Range:
1 to 525 (NTSC, PAL-M); 625 (PAL, SECAM,
SDTV I-576); 750 (HDTV P720); 1125 (HDTV
I1080, HDTV P1080)
Increment: 1
*RST:
1
Manual operation:
See
15.5.5
Pattern trigger
15.5.5.1
Pattern definition
.......................................................................................... 338
.........................................................................................339
....................................................................................... 339
TRIGger:A:PATTern:SOURce
<SourceString>
Sets the state for each channel.
Parameters:
<SourceString>
String containing 0, 1, or X for each channel.
1: high, the signal voltage is higher than the trigger level.
0: low, the signal voltage is lower than the trigger level.
X: Don't care. the channel does not affect the trigger.
Without MSO option R&S
RTB-B1, the pattern has 4 or 2 bits,
depending on the number of channels: <C1><C2>[<C3><C4>].
With MSO option, the pattern has 20 or 18 bits:
<C1><C2>[<C3><C4>]<D0><D1><D2>...<D15>.
Trigger