Remote Control - Commands
R&S NRP
1144.1400.12 6.62
E-3
[SENSe[1..4]]:TRACe:AVERage:COUNt:AUTO:RESolution[?] 1|2|3|4
Defines the number of significant places for the Scope mode for linear units and the number of decimal
places for logarithmic units which should be free of noise in the measurement result. This setting is also
performed by the
DISP[1..4]:RES
command (
Æ
page
6.35
) which tries to set the sensors involved in
the measurement result accordingly. (For further details, see Chapter 4.) However,
SENS[1..4]:AVER:COUN:AUTO:RES
does not affect the DISPlay command. The parameters of the
two commands are different but have the same meaning.
SENS[1..4]:AVER:COUN:AUTO:RES
1 2 3 4
DISP[1..4]:RES
1 0.1
0.01
0.001
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:TRACe:AVERage:COUNt:AUTO:POINt[?] <int_value>
Specifies the measured value in the Scope mode used for the automatic determination of the filter
length.
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:TRACe:AVERage:COUNt:AUTO:NSRatio[?] <float_value>
Indicates the maximum noise ratio in the measurement result for the Scope mode.
The setting is taken
into account only if
SENS[1..4]:TRAC:AVER:COUN:AUTO ON
and
SENS[1..4]:TRAC:AVER:
COUN:AUTO:TYPE NSR
are set.
Unit:
DB | PCT
Default unit:
DB
Value range:
depending on sensor
*RST value
:
depending on sensor