R&S NRP
SCPI Error Messages
1144.1400.12
9.5
E-2
Query Errors
Query errors; cause bit 2 in the ESR register to be set
Error code
Error text returned upon queue query
Explanations
-400
Query error
Generic query error that cannot be defined more precisely.
-410
Query INTERRUPTED
The query was interrupted.
Example: The query is followed by new data before the response was completely sent.
-420
Query UNTERMINATED
An incomplete query was received.
-430
Query DEADLOCKED
The query cannot be processed.
-440
Query UNTERMINATED after indefinite response
A query is received in the same program message after a query requesting an indefinite response.
Device-Dependent Errors
Device-dependent error; causes bit 5 to be set in the ESR register
Error code
Error text returned upon queue query
Explanations
1
Device-dependent error
This error message is sent when the device cannot detect a more specific error.
2
IEEE 1174 error (unknown emulation code)
An invalid IEEE-1174 emulation mode was selected for the Ethernet link.
22
IEEE1174 mode 488.1 not allowed
Setting the IEEE 1174.1 mode is not permissible.
24
Sensor mode not supported
An attempt was made to set the sensor to a measurement mode which is not supported.
25
Not supported
A command is not supported (by the sensor).
26
State not supported
Attempt was made to set an invalid value. The permissible values for sensor-dependent commands can be
declared as impermissible by a sensor depending on the context.
27
Sensor not present
A command could not be executed because the required sensor is not connected.