
R&S NRP
Measuring average power (Cont Av mode)
1144.1400.12 2.7
E-2
5
Setting a fixed offset correction
–
( )
*
↔
*
¾
Using the topmost rocker switch or the cursor keys
(
ÕÖ
), select the
Sensor
menu.
All sensor-related settings can be made in the
Sensor
menu. This determines the type and details of data ac-
quisition.
Offset…
| Filter
–
( )
*
¾
Press the
Offset…
softkey (left-hand side of rocker-
switch).
The Offset dialog box opens. Factors for correcting ex-
ternal signal losses or gains, due to an attenuator, for
example, can be set in this box.
Global
–
( )
*
¾
Activate the editor for the global offset by pressing
the left-hand side of the rocker switch next to
Global
.
Positive values are for correcting losses and negative
values for gains.
1 2 3 4 5 6 7 8 9 0 . –
( )
*
*
¾
If you have an attenuator at hand, enter its value and
connect the attenuator between the sensor and the
signal source, otherwise just enter 10 dB.
(¶ MENU)
*
¾
Confirm the entry.