Remote Control - Commands
R&S NRP
1144.1400.12 6.60
E-3
[SENSe[1..4]][:POWer]:TSLot:COUNt[?] <int_value>
Sets the number of simultaneously measured timeslots in the Timeslot mode.
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]][:POWer]:TSLot:[AVG]:WIDTh[?] <float_value>
Sets the length of the timeslot in the Timeslot mode.
Unit:
S
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:SAMPling[?] FREQ1 | FREQ2
If the signal to be measured has modulation sections just above the video bandwidth of the sensor
used, measurement errors might be caused due to aliasing effects. In this case, the sampling rate of the
sensor can be set to a safe lower value (
FREQ2
). However, the measurement time required to obtain
noise-free results is extended compared to the normal sampling rate (
FREQ1
).
*RST value
:
depending on sensor
[SENSe[1..4]]:SGAMma[:MAGNitude][?] <float_value>
The R&S NRP can perform a measured-value correction taking the complex reflection coefficient
(source gamma) of the signal source into account. SENS[1..4]:SGAM:CORR:STAT ON must be set.
The magnitude of the reflection coefficient is set with SENS[1..4]:SGAM[:MAGN].
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:SGAMma:PHASe[?] <float_value>
The R&S NRP can perform a measured-value correction taking the complex reflection coefficient
(source gamma) of the signal source into account. SENS[1..4]:SGAM:CORR:STAT ON must be set.
The phase angle of the reflection coefficient is set with SENS[1..4]:SGAM:PHAS.
Unit:
-
(Degree is assumed as unit)
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:SGAMma:CORRection:STATe[?] ON | OFF
Switches the measured-value correction of the reflection coefficient effect of the source gamma
ON
or
OFF
.
*RST value
:
depending on sensor