Setting up a measurement
R&S
®
NRPM
50
Mannual 1425.8663.02 ─ 08
7. On the computer, start a software application to execute the measurement,
Chapter 7, "Performing measurements"
6.3.4
Using the LAN connection
Requires the R&S
NRPM3N LAN sensor module with networking capabili-
ties.
Depending on the available equipment, you have several options to connect a
R&S
NRPM3N LAN sensor module to the controller PC.
Setup with a PoE Ethernet switch
To connect the R&S
NRPM3N to a LAN, use a PoE switch, e.g. the R&S
NRP-
ZAP1. This switch provides the power supply and control for the antenna mod-
ules.
Required equipment
●
1 R&S
NRPM3N sensor module
●
2 RJ-45 Ethernet cables,
To avoid any impact, use at least category 6
STP cables for the LAN connec-
●
1 Ethernet switch, supporting PoE power delivery, e.g., R&S
NRP-ZAP1
●
Controller PC with LAN interface and installed VISA driver or R&S
NRP
Toolkit
and R&S
Power
Viewer
Mobile
Setup
Figure 6-7: Setup with a PoE Ethernet switch
Connecting an R&S
NRPM OTA power measurement