Maintenance, storage and disposal
R&S
®
NRPM
212
Mannual 1425.8663.02 ─ 08
14
Maintenance, storage and disposal
The product does not require regular maintenance. It only requires occasional
cleaning. It is however advisable to check the nominal data from time to time.
14.1
Cleaning
To clean the R&S
NRPM3(N) sensor module
1. Disconnect the sensor module from the antenna modules.
2. Disconnect the sensor module from the controller PC.
3. Clean the outside of the R&S
NRPM3(N) sensor module using a lint-free
cloth. You can dampen the cloth with water but keep in mind that the casing is
not waterproof. If you use isopropyl alcohol or ethanol instead of water, be
careful not to damage the labeling.
Do not use cleaning agents that can damage the instrument such as solvents,
acids or alkalis.
4. To dislodge any particels from the antenna interface connector, use clean
compressed air.
To clean the antenna modules
► To dislodge any particels from the waveguide of an antenna module, use
clean compressed air only.
Do not use any liquids for cleaning.
14.2
Storage
Put plastic end caps on the RF connectors to protect them from damage. Protect
the sensor module against dust.
Ensure that the environmental conditions, e.g. temperature range and climatic
load, meet the values specified in the data sheet.
Storage