Network and remote operation
R&S
®
NRPM
102
Mannual 1425.8663.02 ─ 08
Bit No.
Meaning
8
Reference PLL locked state
This bit signals whether the PLL for the clock reference is synchronized. The bit is
useful when selecting an external clock source.
The following states are possible:
●
Internal clock (
[SENSe<Sensor>:]ROSCillator:SOURce INTernal
INT
):
–
1
(always)
●
[SENSe<Sensor>:]ROSCillator:SOURce EXTernal
):
–
1
if the sensor module was able to synchronize with external clock
–
0
if the sensor module could not synchronize with external clock
9 to 15
Not used
9.2.6
Questionable status register
Contains information on questionable sensor module states. Such states may
occur when the sensor module is not operated in compliance with its specifica-
tions.
Querying the register:
●
STATus:QUEStionable:CONDition?
●
Status reporting system