Remote control commands
R&S
®
NRPM
146
Mannual 1425.8663.02 ─ 08
Example: Repeating termination control
Further settings for this example:
●
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol
trol
REPeat
is mandatory in trace mode)
Every chopper phase is started by a trigger event and lasts the defined trace
time. During a chopper phase, the trigger synchronization is high (
). After 2 chopper phases, 1 measurement is completed. As
defined by the trace average count, after 2 measurements, the trace measure-
ment result is averaged and available.
1
2
3
4
5
6
7
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Trace measurement result
6 = Trigger synchronization
7 = Return to the start of the measurement cycle
Controlling the measurement