Operating concepts
R&S
®
NRP Series
43
Getting Started 1419.0170.02 ─ 18
Starting a measurement
For a detailed description of how to measure in this setup, refer to the user man-
ual of the R&S
NRX.
1. Preset the R&S
NRX and the connected R&S
power sensors.
a) Press the [Preset] key.
b) Tap "Preset".
All parameters are set to their defaults.
2.
Note:
Turn off all measurement signals before zeroing. An active measure-
ment signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [Zero] key of the R&S
NRX.
c) Tap "Zero All Sensors".
3. Configure the measurement.
a) In the "Measurement Settings" dialog, select the "Measurement Type", for
example "Continuous Average".
b) Tap "Quick Setup"
> "Auto Set".
4. Switch on the signal source.
The measurement starts, and the result is displayed in dBm.
5. If necessary, perform further settings.
R&S
NRX