Operating concepts
R&S
®
NRP Series
42
Getting Started 1419.0170.02 ─ 18
5.7
R&S
NRX
In a measurement, the R&S
NRX uses all power sensor-dependent measurement
functions and displays the results. Thus, you can configure both the measure-
ment and the power sensor.
Required equipment
●
R&S NRP power sensor
●
R&S
NRP
‑
ZK8 cable to connect the power sensor to the R&S
NRX
●
R&S
NRX
Setup
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
1
2
3
4
6
5
Figure 5-4: Setup with an R&S
NRX base unit
1 = Signal source
2 = R&S NRP power sensor
3 = Host interface connector
4 = R&S
NRP
‑
ZK8 cable
5 = Sensor input connector of the R&S
NRX
6 = R&S
NRX base unit
1.
NOTICE!
Incorrectly connecting or disconnecting the power sensor can dam-
age the power sensor or lead to erroneous results. Ensure that you connect or
disconnect the power sensor as described in
Connect the power sensor to the signal source.
2. Connect the cables as shown in
If the power sensor is a R&S
NRP LAN power sensor, you can set up a LAN
connection instead of using the sensor input connector of the R&S
NRX. See
Chapter 3.7.3, "Using a LAN connection"
R&S
NRX